Troubleshooting

Problem
SpO2 or pulse rate cannot be
shown normally
SpO2 or pulse rateis shown unsta-
bly
The Oximeter cannot be powered
on
Display suddenly powers off.
"Err 3" or "Err 4"
"Err 7"

Symbol Definitions

The following symbols are associated with your Advantage™ 2200
The equipment type is BF
Refer to user manual before use
Hemoglobin saturation
Heart Rate (BPM) Beats per Minute
PR BPM

Limited Warranty

American Diagnostic Corporation (ADC®) warrants its products against defects in materi-
als and workmanship under normal use and service as follows:
1. Warranty service extends to the original retail purchaser only and commences with the
date of delivery.
2. Your Pulse Oximeter is warranted for one (1) year from date of purchase (all parts).
What is Covered: Replacement of parts, and labor.
What is Not Covered: Transportation charges to and from ADC®. Damages caused by
abuse, misuse, accident, or negligence. Incidental, special, or consequential damages.
Possible Reason
1. Finger is not inserted correctly.
2. Oxyhemoglobin value is too low to
be measured
1. Finger might not be inserted deep
enough.
2. Excessive patient movement.
1. No battery or low battery power.
2. Battery not installed correctly.
3. The Oximeter may be damaged.
1. The Oximeter is automatically pow-
ered off, when no signal was detected
after 8 seconds.
2. Low battery strength.
1. Low Power.
2. Mechanical device error near sen-
sor
1. Low Power
2. Oximeter is damaged
Solution
1.
Re-insert/re-position finger
2.
Ensure that sensor is clean and not
blocked
3.
Try taking a reading on a different
finger or warm your finger by rubbing.
1. Re-insert/re-position finger.
2. Sit calmly and retry.
1. Replace battery.
2. Remove and reinstall battery.
3. Contact customer service center.
1. Replace the batteries
2. Re-insert/re-position finger
1. Replace the batteries.
2. Return to Service Center
1. Change batteries.
2. Return to Service Center