Solinst Levelogger Gold User Manual page 35

Levelogger series
Table of Contents

Advertisement

Note:
By selecting
more than one
solution, you will
be performing a multipoint
calibration.
Note:
Ensure that
the calibration
solution covers
the entire sensor cell and the
logger is agitated to release
entrapped air bubbles.
Note:
Always remember
to use fresh
solution, remove
all bubbles from the sensor
pins, and allow for thermal
equilibration of the sensor
before proceeding.
Levelogger User Guide - Software Version 3.4.0 or Higher
2. Setup: The Levelogger Software requires the user to choose the calibration
solutions. You can choose up to three solutions for a multipoint calibration.
Ensure the solutions are ready in the calibration beaker or container. Start
the calibration by selecting Next.
3. LTC Levelogger Junior Rinsing Process:
Levelogger first and then rinse the LTC Levelogger Junior with the displayed
calibration solution. Use fresh solution for calibration, and immerse the LTC
Levelogger Junior in the fresh solution. Lightly tap the LTC Levelogger
Junior to remove any bubbles from the sensor. Allow 2-3 minutes, then
select next to calibrate.
Figure 6-4 LTC Calibration Wizard Step 2
Figure 6-5 LTC Calibration Wizard Step 3
(example of window calibrating with 1413 µS/cm solution)
Use DI water to rinse the
Page 30