Explanations Of Functions; Dip Switch; Torque Settings - HIOS HDP-50 Operation Manual

Digital torque meters
Table of Contents

Advertisement

■ Explanations of functions

DIP switch

Always turn off the power before
changing DIP switch settings.
• Auto-reset
DIP switch settings
Switch number
1 sec
2 sec
Time to auto-
1
reset
5 sec
OFF
Printer ON
Data output
Printer OFF
Notes
1
The reset switch does not function when auto-reset is on.
2
The error display (see below) appears when a printer
or the Mitutoyo Input Tool is connected for data out-
put and the LED lights.
Error display
• Data output
1. Set the 【 MODE 】 switch to "PEAK" or "F.PEAK."
2. Set data output DIP switch number 3 to the on
position.
3. Connect a printer or the Mitutoyo Input Tool
to the Digital data output port. (Pressing the
【 RESET 】 switch in this state will result in the
error display. An error will also result if you try
to print out data with a printer that cannot print
out numerical data containing decimal points.
4. Conduct measurement, then press the 【 RESET 】
switch to output the data. Average ("AVE") val-
ues cannot be output.
Output of data from the memory (memory func-
tion on)
Note: Please cancel auto-reset.
1. Press the reset button for 2 seconds.
2. When the display shows the let-
ters "Prn," press reset.
3. All the torque data, from memory number
0001, will be output. The display flickers
during data transfer.
Setting torque
The digital torque meter is shipped with the warn-
ing tone set to sound when maximum torque load is
reached. This tone also sounds when the torque set-
A.RESET TIME P.ON/OFF
1
2
3
4
ON
ON
ON
OFF
OFF
ON
OFF OFF
2
ON
OFF −
ting has been changed. The torque setting can be
changed as desired. The warning tone signals that
the set torque value has been measured.
Example (torque set at 100)
Data is output only for measurements up 100.
Track
Setting
Warning
ON
tone
Switch functions for setting torque
Switches used
SET
① 【 SET 】 switch
1. Press for 2 seconds to enter torque setting mode.
Press for 2 seconds once again to exit torque set-
ting mode.
2. When the meter is in
torque setting mode,
the 4-digit display
can be adjusted, one
digit at a time. Use the SET switch to select the
digit for adjustment. The selected digit blinks
on and off. Each time the SET switch is pressed,
the next digit to the left is selected (or, if pressed
while the leftmost digit is selected, the rightmost
digit will be selected.)
② 【▲】 switch
When the meter is in torque setting mode, this
switch increases
t h e s e l e c t e d
number by one
each time it is
pressed. Note
that the leftmost
digit can only be
0 or 1.
【 RESET 】 switch
P r e s s i n g t h i s
switch readies
t h e m e t e r t o
begin measure-
ment. Pressing it
while in torque
s e t t i n g m o d e
cancels the pro-
cess and takes
the meter out of
torque setting
mode.
- 6 -
Peak
ON
RESET
Setting torque
0 or 1
0 - 9
0 - 9
Setting torque
(Example: Changing a torque
setting from "0000" to "0100")
Ready for
measurement
Press for 2 seconds
① 【SET】switch
to enter torque
setting mode.
The last 0 is now
selected.
The last 0 flickers
Press twice to
① 【SET】switch
select the third 0.
Press once to
② 【▲】switch
increase the selected
digit from to 1.
Press for 2 seconds
① 【SET】switch
to exit torque
setting mode.
Reset