5. Operation
Each radio probe has a probe ID (RF ID). This consists of
the last 3 digits of the serial no. and the position of the
slide switch in the radio probe (H or L).
Configuration mode is opened and RF ID and Auto are
lit.
The radio probe is switched on.
Select the desired option with
1
:
· YES: Switches automatic probe detection on
(recommended).
· no: Switches automatic probe detection off.
no was selected:
Use
2
confirm with
YES was selected:
- Automatic probe detection is started. Auto flashes
while the instrument looks for a radio probe that is
switched on.
- Once a radio probe is found, the probe ID is
displayed. If no probe is found, NONE lights up.
Possible reasons why probes are not found:
· The radio probe is not switched on or the
· The radio probe is outside the range of the
· Sources of interference are influencing the radio
If necessary, rectify the possible causes for the
disruption to the radio transmission and start
automatic probe detection again with
/
to set the probe ID manually and
battery of the radio probe is spent.
measuring instrument.
transmission (e.g. reinforced concrete, metal
objects, walls or other barriers between
transmitter and receiver, other transmitters of the
same frequency, strong electromagnetic fields).
.
and confirm with
.