Calibration unit

Accessories
Visual check
Function check
Electrical value:
Where?
Desired value:
Figure 5-11
Connection cable 1
5-14
Multimeter, PC with test- and calibration software
The upper, lower and lateral travel paths of the calibration cell slide must be free of
interference by cables and tubes.
• The calibration cell installed in the calibration unit should travel to the lower
position when the analyzer module is turned on (measurement/zero calibration).
• The calibration cells can be moved individually using MASSERV.
Device test menu
Voltage
Plugs X6 or X7 Pin 2 + to pin 1 ⊥
Power supply: +15V
Chapter 5: Testing
SHB_U14