View the calibration data
Data from the most recent calibration can be shown.
1. From the main menu use the
Confirm.
2. Select Display data.
3. Select Calibration data and confirm with
is shown.
• pH-the slope and offset values are shown alternating with the
deviation (in %) and calibration temperature.
• ORP-the measured mV value and calibration temperature are
shown.
• Conductivity-the cell constant and calibration temperature for each
standard are shown.
Set the calibration reminder
The calibration reminder can be set between 0 to 23 hours or 1-7 days
(default 1 day). The display shows the remaining time to the new
calibration.
Note: When 0 days is selected, the calibration reminder is turned off.
1. From the main menu use the
Confirm.
2. Use the
to enter the calibration menu.
3. Use the
or
to select Cal. frequency and confirm.
4. Use the and to advance to the next step and use the
change a value. Confirm.
Push
to start the calibration.
Sample measurements
Each probe has specific preparation steps and procedures for taking
sample measurements.
1. From the main menu use the
MEASURE. Confirm.
10 English
or
to select DATA LOGGER.
. The last calibration data
or
to select CALIBRATION.
or
or
and and to select
2. Use the
to change the following settings. Confirm every entry.
Option
Description
Resolution
Select the resolution: 1, 0.1, 0.01 (default) or 0.001
Measure
Stability—select By stability Criterion: Fast (variation <
0.02 pH in 6 s), Standard (variation < 0.01 pH in 6 s) or Strict
(variation < 0.002 pH in 6 s). In continuous—enter the time
interval for the In continuous Acquis. interval (data storage or
printing data). By time—enter the time interval for data storage
or printing data.
Display mV
Display mV—select YES or NO to show mV.
Limits
Limits—select YES or NO. YES: Enter the upper and lower
limits. An acoustic warning appears when the measurement is
out of limit. The report output shows an A next to the
measured value when the measurement was out of limit.
Isopotential Isopotential—change the Isopotential pH value in Data
introduction. Select Calculate to calibrate the probe again.
3. Push
to start the measurement.
Note: If the measurement is not stabilizing after 120 seconds, the meter turns
automatically into the continuous measurement mode.

Advanced operation

Use a sample ID
The sample ID tag is used to associate readings with a particular sample
to
location. If assigned, stored data will include this ID.
1. From the main menu use the
2. Use the
or
to select Sample ID and confirm.
or
to select SYSTEM. Confirm.