  • Page 1 B100/B200 Series Biological Microscope Operation Instruction To ensure the safety and obtain satisfactory performance, please study this operation instruction thoroughly before your operation.
  • Page 2: Table Of Contents

    Contents 1.Purpose……………………………………………………………………………2 2.Specifications………………………………………………………………………2 3.Installation…………………………………………………………………………4 4.Operation……………………………………………………………………………4 5.Maintenance…………………………………………………………………………8 6.Troubleshooting……………………………………………………………………10 7.Outfits……………………………………………………………………………13...
  • Page 3 Fig.1 B200 Binocular Biological Microscope Monocular Head Eyepiece Main Body Nosepiece Clamp O bjective Condenser Up- down Knob S tage Coaxial Coarse And Fine Focusing Knob Condenser Electric Base Light Controller Power S witch And Light Controller Fig.2 B100 Monocular Biological Microscope...
  • Page 4: Purpose

    1. Purpose B series biological microscope is designed for teaching, clinical verify. The series microscope with original style, steady structure, convenient operation and clear image is suitable for observing various biological specimens, and mostly applied in school, college, hospital and family. 2.
  • Page 5 2.3 The other specification 2.3.1 Mechanical tube length: 160mm 2.3.2 Conjugate distance: 195mm 2.3.3 Head: Monocular, Inclined 30°, Rotable 360°, Eyepiece can be fixed, Anti-fungal systems. Seidentopf binocular head 30°, Rotable 360°, Interpupillary adjustable distance is 55-75mm, Diopter adjustable range ±5, Anti-fungal systems.
  • Page 6: Installation

    Rotation potentiometer with power switch Fuse 2A ф5×20 3. Installation B series microscope is packed integrate, and it is convenient for installation. Open the package carton, and insert eyepiece into the eyepiece tube of the head (eyepiece for monocular microscope has been inserted before packing). If objectives don’t be fixed, please take objectives out from objectives packing box, and drive them into the holes of nosepiece orderly and tightly according to the times.
  • Page 7 Please take the two eyepiece tubes to turn until the bright rings observed by two eyes are in superposition entirely. ▲B100 monocular microscope don’t need to be adjusted. 4.4.2 Diopter adjustment As focusing for binocular, user should observe right eyepiece with right eye, and...
  • Page 8 left eyepiece clear as same as the right eyepiece. 4.5 Focusing adjustment Usually 10X objective should be used for focusing adjustment first. Take 10X objective into bright path, then observe the specimen with right eyepiece by right eye and turn coarse focusing knob to find the image, then turn slowly fine focusing knob to make the image clear.
  • Page 9 of operation. Rotation clockwise makes tension decrease, and by contraries, rotation counterclockwise makes tension increase. 4.7 Usage of immersion oil objective The 100X objective of B series microscope can be used for observation even if without immersion oil. However, adding moderate immersion oil between the front lens of 100X objective and the cover glass of specimen can make the image clearer.
  • Page 10: Maintenance

    Take light collector down from base, then take mirror on base. Turn mirror to make reflecting face to the light source, and make reflecting face and horizontal face as 45° approximately to make the light to specimen as possible. The mirror has two sides, one is a plane mirror, and the other one is a concave one.
  • Page 11 5.1.4 Plastic should be cleaned by soft cloth with clear water. 5.2 Environment of using and placing 5.2.1 Microscope should be used and placed in a cool, dry, non-dust, non-shake and non-corrosive gases environment. 5.2.2 Microscope should be used in environment of indoor temperature 0°-40°C and maximum relative humidity 85%.
  • Page 12: Troubleshooting

    lamp housing cover on the underside of base. 5.3.4 Pull over the lamp housing cover. 5.3.5 Pull out the bulb should be replaced, hold a new bulb with silk cloth to avoid fingerprint and dust affect bulb brightness and service life, and insert fully the contact pins into the bulb socket.
  • Page 13 Trouble Causation Remedy Plug is unreliable Plug in again Bulb is broken Change bulb Switch on but bulb dark Fuse is broken Change fuse Bulb is unstable Insert it again Bulb is flickering or brightness is unsteady Bulb is broken Replacing bulb Bulb specification doesn’t Replacing bulb...
  • Page 14 (contrast or definition Surface of objective lens is Clean it isn’t enough) dirty (especially it is easy for the front lens of 40X objective to dip in immersion oil) Immersion oil isn’t used for Use immersion oil 100X objective (oil) Immersion oil doesn’t meet Use immersion oil the requirement...
  • Page 15: Outfits

    B Series Biological Microscope Outfits Items Specification B102 B103 B104 B202 B203 B204 Order No. ● ● ● ●● ●● ●● WF 10X-18mm WF10 ○ ○ ○ ○○ ○○ ○○ WF 16X-13mm WF16 WF 10X-18mm ○ ○ ○ ○ ○ ○...
  • Page 16 Single Condenser ● ○ (5 Aperture Disc Condenser Diaphragm) Abbe N.A. 1.25 ● ● ● ● ● (Iris Diaphragm) Dark Field Available For ○ ○ ○ ○ ○ Ring 4X-40X Objective Mirror ○ ○ ○ ○ ○ ○ Built In Blue Filter Incandescent Lamp 110V/220V, 20W ●...