Calibration - Eiki EK-610U Service Manual

Hide thumbs Also See for EK-610U:
Table of Contents

Advertisement

4-4 Calibration

1. Lens Calibration
Note: Always perform a Lens Calibration after repairing the projector
or replacing the light engine.
Inspection Item - Check if the Lens Shift Module is moving smoothly.
Criteria
Procedure
2. G Sensor Calibration
Note: Always perform a G Sensor Calibration after repairing the
projector or replacing the main board.
Procedure
- If the Lens Shift Module is not moving smoothly
or sound noising,please replace the light engine module.
- Place the projector on a clean horizontal surface.
- On the Remote, press Menu >Lens Function.
- Select Lens Calibration.
- Press Enter.
- Place the projector on a clean horizontal surface.
- Get into service mode seclect "G seneor
calibration".
- Press "Enter" to excute G sensor calibration.
EK-610U/EK-611W/EK-612X
4-4

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ek-611wEk-612xEk-610 series

Table of Contents