Safety 
The MN62 is designed in compliance with standards such as IEC/EN61010‐1, pollution grade II, 
overvoltage CAT III 600V and double insulation standards. Please comply with the instructions 
provided in this guide otherwise the protection provided by the instrument may be affected. 
WARNING
CAUTION
MAX
600V
 
PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY 
OVERVOLTAGE CATEGORY I 
Equipment  of  OVERVOLTAGE  CATEGORY  I  is  equipment  for  connection  to  circuits  in  which 
measures are taken to limit the transient overvoltages to an appropriate low level. 
  Note – Examples include protected electronic circuits. 
OVERVOLTAGE CATEGORY II 
Equipment  of  OVERVOLTAGE  CATEGORY  II  is  energy‐consuming  equipment  to  be  supplied  from 
the fixed installation. 
  Note – Examples include household, office, and laboratory appliances. 
OVERVOLTAGE CATEGORY III 
Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations. 
  Note – Examples include switches in the fixed installation and some equipment for industrial use 
with permanent connection to the fixed installation. 
OVERVOLTAGE CATEGORY IV 
Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the installation. 
Note – Examples include electricity meters and primary over‐current protection equipment 
This symbol adjacent to another symbol, terminal or operating device indicates 
that the operator must refer to an explanation in the Operating Instructions to 
avoid personal injury or damage to the meter. 
This WARNING symbol indicates a potentially hazardous situation, which if not 
avoided, could result in death or serious injury. 
This CAUTION symbol indicates a potentially hazardous situation, which if not 
avoided, may result damage to the product. 
This symbol advises the user that the terminal(s) so marked must not be 
connected to a circuit point at which the voltage with respect to earth ground 
exceeds (in this case) 600V AC rms or DC. 
This symbol adjacent to one or more terminals identifies them as being 
associated with ranges that may, in normal use, be subjected to particularly 
hazardous voltages. For maximum safety, the meter and its test leads should 
not be handled when these terminals are energized. 
This symbol indicates that a device is protected throughout by double 
insulation or reinforced insulation. 
3
 MN62‐en‐GB_V1.1   12/15 