Entering Pulse Number

Procedure Options
1-14
5. Enter the number of pulses desired (between 1 and 10,000) and
[ENT]
press the
key. That number of pulses will then be output at
the specified frequency.
6. Repeat Steps 3, 4 and 5 to test all four of the frequency/totalizer
outputs. If any of the outputs fails to pass the test, contact the
factory for assistance.
You have completed testing the totalizer/frequency outputs. Do one of
the following:
To calibrate/test additional slots, press
the appropriate section.
To leave the Calibration Menu, press
[F4] (EXIT)
and proceed to
[F4] (EXIT)
twice.