E16: High Signal

Screen Messages

Message
Backup Battery FAIL
Backup Battery OK
EPROM sum = XXXX
FRIGID_INIT Executed
NVR FAIL
NVR OK
All Logs Cleared! hit key
All Sites Cleared! hit key
Do you want to SAVE?
Error Codes and Screen Messages
Problem:
The meter signal strength exceeds the limits entered via
the User Program.
Cause:
A signal that exceeds the programmed limits is probably
caused by the entry of an improper value in the
TRANSDUCER
Action:
Check the values programmed into the
PIPE
submenus, as described in the Startup Guide and the
Programming Manual. Bypass or remove the preamp.
A variety of messages may appear on the display screen during the
performance of a task. Since the error codes have already been
discussed in this chapter and the locator bar messages are discussed in
detail in Chapter 3, Operation, of the Startup Guide, they will not be
repeated here. All other messages are listed in Table 2-1 below.
Table 2-1: Screen Messages
Power Up Messages
The backup battery that powers the non-volatile RAM has failed. Contact
the factory.
The backup battery that powers the non-volatile RAM has passed.
Record the EPROM sum at the initial power up and periodically thereafter.
The NVR (non-volatile RAM) was automatically initialized due to a memory
fault. Contact the factory.
The non-volatile RAM has failed. Contact the factory.
The non-volatile RAM has passed.
Measurement Mode Messages
This message may appear during one of the following tasks:
1. clearing a log - there are no more logs to clear
2. logging data - user has hit [ENT] instead of selecting an old log or
entering a new name
3. printing a log - there are no logs to print
4. displaying a log - there are no logs to display
Hit any key to resume taking measurements.
This message may appear during one of the following tasks:
1. saving a new site - a new site name was not entered or overwrite
existing site was not selected
2. recalling a site - there are no site files to recall
3. clearing a site - there are no sites to delete
4. printing a site - there are no sites to print
5. displaying a site - there are no sites to display
Hit any key to resume taking measurements.
This message appears upon exiting the User Program, if the new site
data has not been saved. Failure to save will result in loss of the new data
next time the site data is changed or recalled.
PIPE
or
submenu of the User Program.
Meaning
April 2008
TRANSDUCER and
2-5