GE DigitalFlow GC868 Service Manual page 30

Parametrics gas clamp-on ultrasonic flowmeter
Hide thumbs Also See for DigitalFlow GC868:
Table of Contents

Advertisement

April 2008
Message
Duplicate name,
Enter another.
End Time must exceed
Start Time by 5 min.
Header invalid
Log Active, END only
hit any key
Log DONE, to inspect
hit any key
Log hasn't started! hit key
OK, settling
Outside limits, value rejected.
Overflow
??P<L Enter L again.
range is X.XX to X.XX
Review calibration
Review parameters
Starting time must exceed
current time
Write error.
2-6
Table 2-1: Screen Messages (Continued)
The site file or log name is already in use. Enter a different name.
This message appears when in the LOG menu. Enter an end time that is
at least five minutes later than the start time.
An option card error indicating a programming failure or a loss of
memory. Contact the factory.
When in the LOG menu, this message indicates that the log is still
compiling data. Only the End Time may be edited.
When in the LOG menu, this message indicates that the log is complete.
Hit any key to display the log.
When in the LOG menu, this message indicates that the log has not yet
started.
The measurement has not stabilized. Wait before taking a reading.
When in the CAL menu, this message indicates that the calibration of the
analog output is invalid. Hit any key to clear the message, and the GC868
will default to the last valid calibration. The message will also appear if
there is no ammeter connected to the analog output during calibration.
The display value overflowed. Reduce the number of decimal digits or
change the units.
The entered path length (P) is less than the axial dimension (L). Enter a
new value for L.
When in the User Program, this message indicates that the entered
number is unacceptable. Hit any key and enter a number within the
allowable range.
An option card error indicating a programming failure or a loss of
memory. Contact the factory.
An option card error indicating a programming failure or a loss of
memory. Contact the factory.
This message appears when in the LOG menu STD option. Enter a start
time that is later than the current time.
An option card error indicating a programming failure or a loss of
memory. Contact the factory.
Meaning