Troubleshooting When Fault Occurs In Viscometer - Hitachi IJ RX2 Service Manual

Hide thumbs Also See for IJ RX2:
Table of Contents

Advertisement

4.9 Troubleshooting when Fault Occurs in Viscometer

Perform the following when abnormality occurs in the viscometer:
Caution: If abnormality occurred when "Viscometer Calibration" was executed
during installation, no "coefficient" can be determined.
If "coefficient" has not been determined, the IJP will automatically
switch to the dot count system and operate.
The references for normal values are 1-3 seconds for fall time, and
1000-2000 for coefficient.
(1) Viscosity Reading Instability
Phenomenon: Some unevenness occurred in 5-time measurement results of fall
Estimated cause: Entrance of foreign object
Troubleshooting:(a) Recalibrate: Proceed with (b) if no recovery is made.
(2) Viscosity Readings Out of Range
Estimated cause: Ink adhered to the IN-side of viscometer.
Troubleshooting: Disassemble and wash viscometer.
Estimated cause: Proximity sensor was faulty.
Troubleshooting: Replace proximity sensor.
If there is no recovery with the above troubleshooting, replace the viscometer.
(3) Viscometer Temperature Sensor Fault
Cause: Thermistor was faulty.
Troubleshooting: Confirm the connection of connector and replace the thermistor.
4.9.1 Viscometer removal procedure
Caution: Disposing of ink is not necessary.
Viscometer
Main tank
1 Removing the viscometer IN-side tube
(A) Pull the main filter and intensifier tank out forward so that the viscometer is
visible.
(B) Release the viscometer IN-side tube (mark G) from clamp.
(C) Turn the lock nut to remove it.
time. (The difference between the maximum and minimum values
of fall time exceeded 10% of average value.)
(b) Disassemble and clean viscometer: Proceed with (c) if no
recovery is made.
(c) Replace viscometer.
Grommet
Lock nut
2 viscometer setscrews
Viscometer IN-side tube (mark G)
4-22
Clamp

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents