3 Wire RTD

Optional PCN 5447 or 5416 board
needed. This board will accept 2
wire RTDs as well. RTD devices
are precision resistors whose
resistance value varies with
temperature. The connection of a
third wire eliminates the natural
resistance of the lead wires to
improve sensor accuracy. The
DQ15D control measures the RTD
resistance (and the third wire
resistance) and compares that
measurement with a standard set
of values stored in the memory.
You can restore, update or verify that
the standard set of values is correct.
Note: For a 1000 ohm sensor, the DIP
switches should be OFF, OFF, OFF.
For 100 ohm, ensure the DIP switches
are ON, OFF, OFF. See Dip Switch
Settings in Configuration (Setup).
Equipment needed
• Two precision resistors (tolerance
+/- 0.1% or better) with a fixed
value equal to the RTD nominal
value (i.e. 1000 or 100 ohms).
• A suitable jumper cable to
facilitate changing input
resistance.
• A short piece of jumper wire
(simulates third wire).
Warning!
Calibration procedures require the
removal of the rear cover of the
control. It also requires that power is
ON, exposing the technician to
potentially lethal voltages. Always
Exercise EXTREME CARE and wear
tested electrician's gloves when
power is on.
Calibration Procedure
Turn OFF all power.
1
Remove rear cover.
2
Remove RTD sensor.
3
M-34-01
Revision - Date: 01 - 04/23/08
Install the short piece of jumper
4
wire from terminal #1 to #3.
Install the precision resistors in
5
place of the RTD sensor, as shown
in terminals #2 and #3.
Precision Resistors in Place
Install the jumper cable between
6
the loose end of one of the resistors
and the fixed end of the other resistor
to establish an input value of a single
resistor (i.e. 1000 ohms or 100 ohms),
as shown.
Establishing Resistor Input Value
Carefully restore power to the
7
controller. Do not come in contact
with any exposed voltage.
Press
and
8
and hold for approximately 6 sec.
The screen displays
is flashing, use
22. Press
.
SET
CAL1
Press and hold
9
The screen displays
message to change to
11
Calibration
simultaneously
While the 0
.
AC.0
to change this to
displays.
for one sec.
SET
Wait for the
.
Hold
.
CAL2