Assembly Replacement and Post-Repair Procedures

Removing the Front Panel Interface and Keypad Assemblies

Removing the Front Panel Interface and Keypad
Assemblies

Tools Required

• T-10 TORX screwdriver
• T-15 TORX screwdriver
• small slot screwdriver
• ESD (electrostatic discharge) grounding wrist strap
• 5/16-inch open-end torque wrench (set to 10 in-lb)

Removal

1. Remove the front panel assembly from the analyzer, refer to
Assembly" on page
2. Remove the 4 ribbon cables (item 1 through 4) from the front panel interface board.
3. Remove the 4 screws (item 5), that attach the interface board.
If the front panel interface board is to be replaced, install the new board at
NOTE
this point and reverse steps 1 through 3 to complete the process.
4. Remove the 7 screws (item 6) from the A1 keypad board.
5. Remove the 3 screws (item 7) from the display cover.
6. Lift the display assembly slightly and slide out the A1 keypad board.

Replacement

1. Reverse the order of the removal procedure.