Transfer Function; Advanced - Philips MML1801 Series User Manual

Lcd 18” monochrome monitor
Table of Contents

Advertisement

6.3.5 ABC Selection
1) Press Enter button when the ABC Selection is highlighted on Image Enhancement
The ABC Selection window appears.
2) Press the Adjust + or Adjust - button until the desired item is highlighted.
- ABC off
Manual back light control regulated by panel button from 250 cd/m² to 500 cd/m²
- ABC full
Automatic control of back light, the back light is function of ambient light and follow a semilogaritmic
courve from 200 cd/m² (at 1 lux) to 500 cd/m² (at 1000 lux).
-ABC reduced
The same as ABC full but with full scale of 350 cd/m² (at 1000 lux).
3) Press the Enter button to confi rm your selection and return to Image Enhancement window

6.3.6 Transfer Function

1) Press Enter button when the Transfer Functiopn is highlighted on Image Enhancement
The Transfer Function window appears.
2) Press the Adjust + or Adjust - button until the desired item is highlighted.
Indicate the output curve relative to the input signal:
- TG3 - PMS
Fixed output curve PMS like.
- Linear
The Tranfer Function is transparent
- DICOM
The output curve follows the DICOM curve
- Monitor TG3
The output curve is the same as PMS standard monitor.
3) Press the Enter button to confi rm your selection and return to Image Enhancement window

6.3.7 Advanced

1) Press Enter button when the Adavanced is highlighted on Main Menu
The Advanced window appears.
ABC Selection
ABC Off
ABC Reduced
ABC Full
Transfer Function
TG3 - PMS
Linear
DICOM
Monitor TG3
19