Wireless Communications - ABB aquamaster Quick Reference Manual

Electronic commercial water meter with integral logger
Hide thumbs Also See for aquamaster:
Table of Contents

Advertisement

...WIRELESS COMMUNICATIONS

Status
GSM Module Status
GSM Module Status [>368] shows the current
GSM radio module status as one of:
0
=
Not Configured
1
=
Off
2
=
Off (SIM not fitted)
3
=
Ready
4
=
Waiting PIN
5
=
Waiting PUK
6
=
Waiting Antitheft PIN
7
=
Waiting Antitheft PUK
8
=
Waiting PIN2
9
=
Waiting PUK2
10 =
SMS Mode
11 =
Command Handling
12 =
Connecting
13 =
Connected
Signal Strength (current value)
The Signal Strength parameter [>348] may be
used to obtain the radio signal strength on
demand.
e.g. >348<0>=16
Signal Log (Signal Strength History)
A signal strength reading is taken each time the
module is powered up (or daily for a mains
powered unit). The last 7 signal strength readings
can be examined using Signal Log in Menu 12
[>365]. The most recent reading is to the left.
i.e. 16 is the most recent reading in this example:
<0>365=16 14 14 13 12 14 14
Total Connect Time
Accumulated connection (call) time can be
obtained using the Total Connect Time parameter
in Menu 12 [>349]. It is given in hh:mm:ss and
excludes SMS text activity.
Access via SMS Text Message
A transmitter with the GSM option also provides a
means of accessing AquaMaster data via SMS
text messaging. An SMS Request Text can be sent
from a mobile phone to AquaMaster and an SMS
Reply Text with the requested information will be
sent to the originating phone or SMS gateway. If
12
the AquaMaster is battery powered, the SMS
Request will not be serviced until the next
programmed WakeUp time as defined in Menu 13.
SMS Request Message
SMS Request messages must have the following
format:
+password;command;command;...;command;
Where:
+password is the character + followed by the
AquaMaster login password [>248] and the
command may be:
Either:
Any of the AquaMaster Parameter Access
Commands:
FLW
Rate Of Flow
VEL
Rate Of Velocity
PRS
Pressure
ALM
Alarm
TOF
Total Volume Forward
TOR
Total Volume Reverse
TON
Total Volume Net
TFA
Tariff A
TFB
Tariff B
TIM
Time and Date
or
any regular Command Line Interface requests
e.g.: >365
to show the last 7 Signal Strength readings.
Example
Sending:
+user;FLW;PRS;TOF;TFA;TIM;>365;
results in a reply similar to:
-AquaMaster;ABB01M;
Pressure=-0.619765 Bar; TOT Fwd=16853 m3;
TRF A=1866 m3; TIME=00:00:01
03;<0>365=14 14 14 13 12 14 14;
Flow=-157.93
l/s;
08-07-

Advertisement

Table of Contents
loading

Table of Contents