SECTION 10—MAINTENANCE

Cleaning/Replacing the Cabinet Filter

Unplug the compressor when cleaning.
DO NOT operate the compressor without the filter installed.
NOTE: For this procedure, refer to FIGURE 10.1.
NOTE: All compressors have one filter on the back of the cabinet.
NOTE: Perform this procedure at least once a week.
1. Pull out the filter.
2. Clean the filter with a vacuum cleaner or wash in warm soapy water and rinse
thoroughly.
3. Dry the filter thoroughly before reinstallation.
Cabinet Filter
FIGURE 10.1 Cleaning/Replacing the Cabinet Filter

Cleaning Cabinet

Unplug the compressor when cleaning. To avoid electrical shock, DO NOT remove
cabinet.
1. Clean the cabinet with a mild household cleaner and non-abrasive cloth or sponge.