SECTION 10—MAINTENANCE

Verification Process - Invacare HomeFill II Compressor

NOTE: Inspection to be performed upon delivery of unit from Invacare and upon return of unit
from rental periods
.
MODEL No.________________________
ON EACH INSPECTION
Record Date of Service
Record Elapsed Hours on the
Hour Meter
Clean and disinfect the
equipment per Cleaning/
Disinfection policy
VISUAL INSPECTION
Check the electrical cord for
damage
Inspect the cabinet for damage
Inspect the oxygen intake
connection for damage, and/or
the presence of oil or grease
Inspect the cylinder fillport for
damage, and/or the presence
of oil or grease
Inspect that the three warning
labels are present
Clean/Replace the filter
(located in back of cabinet –
IVC Part # 2000489)
NOTE: If any damage is found, return unit to Invacare. If any of the warning labels are missing,
please order replacement labels from Invacare.
Function Test (procedure
below)

Function Test

1. Set test concentrator to 3 L/Min. and verify oxygen concentration > 90%.
2. Plug the unit in.
3. Connect the concentrator to the compressor.
4. Connect an empty HomeFill II cylinder to the cylinder fillport.
5. Turn the compressor on and observe for the following:
A. The Wait (YELLOW) light comes on for 3 minutes.
B. After 3 minutes, the Fill (GREEN) light will come on when the O
concentrator is > 90%.
HomeFill™II Compressor
SERIAL No._______________________
56
from the
2
Part No 1141487