Replacing Hour Meter; Replacing O Inlet Connector/Inlet Connector Dust Cap - Invacare HomeFill II Service Manual

Hide thumbs Also See for HomeFill II:
Table of Contents

Advertisement

SECTION 7—BASE COMPONENTS

Replacing Hour Meter

NOTE: For this procedure, refer to FIGURE 7.3 on page 34.
1. Remove the cabinet. Refer to Removing/Installing the Cabinet on page 21.
2. Note the orientation of the two wire connectors before removing.
3. Remove the two wire connectors from the hour meter.
NOTE: The hour meter and retaining clip are separate components. Both of these components hold
the other component in place. When the hour meter is removed, in STEP 3, the retaining clip will
become a loose component.
4. Pull the retaining clip away from the sides of the hour meter while pushing out the
hour meter from the inside of the base.
5. Position the retaining clip on the inside of the compressor base assembly.
6. Insert the hour meter through the front of the base assembly and through the retaining
clip.
7. Pull the retaining clip forward toward the base assembly so that the hour meter and
retaining clip are snug against the base assembly.
8. Install the two wire connectors onto the hour metere in the orientation previously
noted.
9. Install the cabinet. Refer to Removing/Installing the Cabinet on page 21.
Connector
Retaining
Clip
Hour Meter
Replacing O
2
Cap
NOTE: For this procedure, refer to FIGURE 7.4 on page 35.
Tools required for this procedure are:
• 5/8 inch wrench
• side cutters
1. Remove the cabinet. Refer to Removing/Installing the Cabinet on page 21.
HomeFill™II Compressor
Wire
FIGURE 7.3 Replacing Hour Meter
Inlet Connector/Inlet Connector Dust
Wire Connector
34
Part No 1141487