Replacing Burst Disk Fitting - Invacare HomeFill II Service Manual

Hide thumbs Also See for HomeFill II:
Table of Contents

Advertisement

Gloves must be worn during the re-installation process to prevent oil from skin
contacting the high-pressure switch.
7. Using a 1-1/8" crows foot, install the high- pressure switch. Torque the high pressure
switch to 50 -55 inch pounds.
8. Install the two wires from the high- pressure switch.
9. Perform Leak Test procedure. Refer to Leak Test on page 49
10. Perform Fill Test procedure. Refer to Cylinder Fill Test on page 50.
11. Install the cabinet. Refer to Removing/Installing the Cabinet on page 21.

Replacing Burst Disk Fitting

Tools Required:
9/16 inch Wrench
1. Remove the cabinet. Refer to Removing/Installing the Cabinet on page 21.
While performing STEP 2, gloves MUST be worn during to prevent oil from skin
contacting the burst disk fitting. Otherwise, damage to the fitting may occur.
2. Using a 9/16" wrench, remove the burst disk fitting from the stage 5 cylinder head.
3. Using a 9/16" wrench, install new burst disk fitting by threading fitting into the
cylinder head. Torque to 65-70 inch pounds.
4. Perform Leak Test procedure. Refer to Leak Test on page 49
5. Perform Fill Test procedure. Refer to Cylinder Fill Test on page 50.
6. Install the cabinet. Refer to Removing/Installing the Cabinet on page 21.
Part No 1141487