System-Level Rise And Fall Time Performance Test - Agilent Technologies N1921A Operating And Service Manual

P-series wideband power sensor
Hide thumbs Also See for N1921A:
Table of Contents

Advertisement

Figure 2-3

System-Level Rise and Fall Time Performance Test

The rise and fall time performance of the instrument path must be quantified
accurately. This test however, is more of a system-level verification, validating the
rise and fall time with the N1911/2A using an actual RF pulse.
Test equipment
Signal generator (N5182A)
Power meter (N1911/2A)
Diode detector
Oscilloscope with bandwidth >1 GHz
Trigger cable (U2032A)
Wideband power sensor (N1921/2A)
System specification
Rise/fall time: ≤13 ns ±% error
Overshoot: ≤5%
Measured rise time percentage error versus signal under test rise time
Test procedure
1 Turn on the N5182A, oscilloscope, and N1911/2A.
2 Allow the system to warm up for approximately an hour before starting the
measurement.
19

Advertisement

Table of Contents
loading

This manual is also suitable for:

N1922a

Table of Contents