Verifying probe input resistance

Specification: 1.0 MΩ ±2%
Equipment Required
Equipment
Digital Multimeter
(DMM)
Connect the DMM probes between the probe tip (ChA) and the ground
1
at the tip of the probe.
Set up the DMM to measure resistance.
2
Record resistance and repeat steps 1 through 3 for ChB.
3
Enter these value in the Test Card.
4

Verifying probe rise time

Specification: ≤ 470 ps
Equipment Required
Equipment
Signal Generator
Oscilloscope
Terminations
Coaxial Cables
Adapters
Adapter
Adapter
C A U T I O N
Allow the probe to warm up for at least 15 minutes prior to starting the
calibration tests. Failure to allow the probe to warm up may cause the probe
to fail tests.
1155A 2-Channel, Low-Mass Active Oscilloscope Probe
Critical Specification
Resistance ±1%
The resistance should read 1.0 MΩ ±20 kΩ.
ChA input resistance ____________.
ChB input resistance ____________.
Qty Critical Specification
1
Square wave, <200 ps, 0 V to 1 Vp,
<5% flatness, period 50 ms to
50 µs.
1
>1.5 GHz bandwidth, 50-Ω source54845A (or equivalent)
50 Ω BNC plug to BNC socket
2
50 Ω, low loss
3
3
BNC to SMA
2
BNC to probe tip
2
BNC to BNC
Verifying probe input resistance
Recommended Model/Part
34401A
Recommended Model or
Agilent Part Number
8131A (or equivalent)
10100C (or equivalent)
10503A (or equivalent)
1250-1200
5063-2174
1250-0080
21