Performance Test

Standing Wave Ratio (SWR) and Reflection Coefficient (Rho)

Performance Test

This section does not establish preset SWR test procedures since there are
several test methods and different equipment available for testing the SWR
or reflection coefficient. Therefore, the actual accuracy of the test
equipment must be accounted for when measuring against instrument
specifications to determine a pass or fail condition. The test system used
must not exceed the system Rho uncertainties shown in the following
tables when testing the U2000 Series USB power sensors.

Table 4-1 Power Sensor SWR and Reflection Coefficient for the U2000A

Table 4-2 Power Sensor SWR and Reflection Coefficient for the U2001A

U2000 Series Operating and Service Guide
Frequency
10 MHz to 30 MHz
30 MHz to 2 GHz
2 GHz to 14 GHz
14 GHz to 16 GHz
16 GHz to 18 GHz
Frequency
10 MHz to 30 MHz
30 MHz to 2 GHz
2 GHz to 6 GHz
Actual
Maximum
Measurement
Rho
Actual
Maximum
Measurement
Rho
Service
SWR
0.070
1.15
0.061
1.13
0.087
1.19
0.099
1.22
0.115
1.26
SWR
0.070
1.15
0.061
1.13
0.087
1.19