Verification; Introduction; Test Equipment Required; Test Equipment - Agilent Technologies 6060B Service Manual

300 watt single input electronic loads
Table of Contents

Advertisement

Verification

Introduction

This chapter contains test procedures that check the operation and calibration of the Agilent 6060B and 6063B Electronic
Loads. The tests are performed from the front panel and can be used to determine which circuits are faulty when
troubleshooting. There are some transient, trigger, and pulse functions that require a GP-IB controller and will not be
verified with manual testing from the front panel. The following tests will verify, with a high level of confidence, that the
Electronic Load is operating properly without testing all of its capabilities.
At the end of this chapter are performance record tables where actual measured values can be recorded.

Test Equipment Required

Table 2-1 lists the test equipment required to perform the tests in this chapter. Test setups for the tests are shown in Figures
2-1 through 2-3. Make sure the sense switch on the rear of the load is set to the LCL position since local sensing is used in
all of the test setups. Use adequate wire gauge when making connections (see Chapter 3 in the Operating Manual).
Note
The Electronic Load must pass the selftest at power turn-on before the following tests can be performed.
If the unit fails selftest, refer to the overall troubleshooting procedures in Figure 3-1 in Chapter 3.
Type
120V/60A Source
Current Monitor
Resistor
Current Monitor
Resistor
Digital Voltmeter
Current Probe with
Amplifier and Power
Supply
Oscilloscope
Table 2-1. Test Equipment Required for Verification
Required Characteristics
0 to 20V/0 to 120A
0 to 60 V/0 to 50A
0 to 500 V/0 to 5A
0.10 ohms @ 15A
0.04% @ 25W
0.010 ohms @ 100A
0.04% @ 100W
dc accuracy of 0.01%
6 digit readout
Sensitivity of 1mA/10 mV to 50MHz with
less than 300µA of noise to 5MHz.
Sensitivity: 1mV
Bandwidth: 20MHz
Recommended Model
Agilent 6031A or equivalent
Agilent 6032A or equivalent
Agilent 6035A or equivalent
Guideline 9230/15
Guideline 9230/100
Agilent 3455A, 3456A, or 3458A
Tektronix A6302 probe, AM503 probe amplifier,
and TM501 probe power supply.
Agilent 54504
2
9

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

6063b