Chapter 4 Calibration Procedures
Relay Verification
(Optional)
34902A Relay Contact Resistance Verification
This optional procedure uses an external DMM to make 4-wire ohms
measurements across the relay contacts on the 34902A. The measured
resistance is the series resistance of the two relay contacts (both
contacts are in the same relay).
Note: Be sure to read "Plug-in Module Test Considerations" on page 87.
See the diagram on page 97 for the required connections for each test
Tests 1 - 4:
(be sure to probe the components at the indicated location). For these
measurements, the 34902A is not installed in the 34970A. Record the
4-wire ohms measurements from the external DMM in the table below.
Note: The connections to the external DMM are different for each of
Tests 1, 2, and 3. Be sure to verify the connections shown in the table
below for each of the three tests.
96
External DMM Ohmmeter Connections
Test
HI
LO
#
1
L300
J101, C12
2
L301
J101, C14
3
L302
J101, C15
4
Add (Test 2 + Test 3)
HI
LO
Sense
Sense
L300
J101, C12
__________ Ohms
L301
J101, C14
__________ Ohms
L302
J101, C15
__________ Ohms
__________ Ohms
Measured
Value