Agilent Technologies 34970A Service Manual page 95

Data acquisition/switch unit
Hide thumbs Also See for 34970A:
Table of Contents

Advertisement

Chapter 4 Calibration Procedures
Relay Verification
... Continued from previous page
External DMM Ohmmeter Connections
Test
Channel
#
Closed*
18
Ch 8
19
Ch 9
20
Ch 10
21
Ch 11
22
Ch 12
23
Ch 13
24
Ch 14
25
Ch 15
26
Ch 16
27
Ch 17
28
Ch 18
29
Ch 19
30
Ch 21
31
Subtract (Test 30 – Test 4)
32
Ch 22
33
Subtract (Test 32 – Test 5)
* Only the channel currently under test should be closed at one time. All other channels should be open.
HI
HI
LO
Sense
P2
P2
P1
P2
P2
P1
P2
P2
P1
P2
P2
P1
P2
P2
P1
P2
P2
P1
P2
P2
P1
P2
P2
P1
P2
P2
P1
P2
P2
P1
P2
P2
P1
P2
P2
P1
P4
P4
P5
P4
P4
P5
LO
Measured Value
Sense
P1
__________ Ohms
P1
__________ Ohms
P1
__________ Ohms
P1
__________ Ohms
P1
__________ Ohms
P1
__________ Ohms
P1
__________ Ohms
P1
__________ Ohms
P1
__________ Ohms
P1
__________ Ohms
P1
__________ Ohms
P1
__________ Ohms
P5
__________ Ohms
__________ Ohms
P5
__________ Ohms
__________ Ohms
Test
Relay
Limit
Measured
K408
2.00Ω
2.00Ω
K409
K410
2.00Ω
K411
2.00Ω
K412
2.00Ω
K413
2.00Ω
4
K414
2.00Ω
K415
2.00Ω
K416
2.00Ω
K417
2.00Ω
K418
2.00Ω
K419
2.00Ω
K522
2.00Ω
K522
2.00Ω
93