Advertisement

Chapter 4 Calibration Procedures

34907A Analog Output

34907A Analog Output

Analog Output Verification Test

This procedure is used to check the calibration of the analog outputs
on the 34907A Multifunction Module. Install the module in slot 200.
Verification checks are performed only for those output values with
unique calibration constants.
Make connections to analog output channels as shown below.
1
34907A
For each analog output, set each output value in the table below.
2
Compare measurement results to the appropriate test limits shown
in the table.
0.000 V
10.000 V
114
To DMM
DAC
Quick
Measured
Output
Check
Output
0.000 V
10.000 V
Q
Error from Nominal
1 year
± 20 mV
± 45 mV
To
DMM