Agilent Technologies 34970A Service Manual page 81

Data acquisition/switch unit
Hide thumbs Also See for 34970A:
Table of Contents

Advertisement

Chapter 4 Calibration Procedures
Internal DMM Verification Tests
AC CURRENT Gain Verification Test
Configuration:
AC Current
LF 3 HZ:SLOW (in the Advanced menu)
Make sure you have read "Test Considerations" on page 71.
1
Select Channel 221, set the AC CURRENT function and the 3 Hz
2
input filter. With the slow filter selected, each measurement takes
7 seconds to complete. Before executing each test, you must press
to enable reading monitoring on the selected channel (or use the
ROUTe:MON command from the remote interface).
Select each range in the order shown below. Compare measurement
3
results to the appropriate test limits shown in the table. (Be certain
to allow for appropriate source settling.)
Input
Current
Frequency
[1]
10 mA
1 kHz
[1]
100 mA
1 kHz
10 mA
1 kHz
[1]
1A
1 kHz
[1] Verify only, no adjustment.
Frequency Gain Verification Test
Configuration:
Frequency
6
Make sure you have read "Test Considerations" on page 71.
1
Select Channel 210, select the FREQUENCY function and set 6
2
Select each range in the order shown below. Compare measurement
3
results to the appropriate test limits shown in the table. (Be certain to
allow for appropriate source settling.)
Input
Voltage
Frequency
[1]
10 mV
100 Hz
1 V
100 kHz
[1] Verify only, No adjustment. For this test, isolate the calibrator's output from earth ground.
Q: Quick performance verification test points.
Quick
Range
check
10 mA
100 mA
Q
1 A
1A
1
digits
2
Quick
Range
Check
100 mV
1 V
Q
Error from Nominal
24 hour
90 day
± 14 µA
± 14 µA
± 14 µA
± 600 µA
± 600 µA
± 600 µA
± 1.41 mA
± 1.41 mA
± 1.41 mA
± 1.4 mA
± 1.4 mA
± 1.4 mA
1
digits.
2
Error from Nominal
24 hour
90 day
0.06 Hz
0 .1 Hz
±
±
±
6 Hz
10 Hz
±
±
±
1 year
4
1 year
0 .1 Hz
10 Hz
79

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents