Chapter 4 Calibration Procedures
Relay Verification
Make the connections to the 34908A as shown in the diagram below.
Tests 4 - 43:
Connect copper shorts between all channels as shown. Be sure to
route your wiring for proper strain relief and install the module cover.
Install the 34908A in slot 200 of the 34970A.
Connections for 34908A Verification Tests 4 through 43
For each test shown in the table starting on the next page, close only
the channels shown in the "Channels Closed" column (closing one
channel will open the previously closed channel). Record the 4-wire
ohms measurements from the external DMM in the table.
4
109