Chapter 4 Calibration Procedures
–10 Vdc Adjustment Procedure (Optional)
Continued from previous page...
Reverse the cable connections to the calibrator to create a -10 Vdc
1
voltage standard. You must physically reverse the cables.
DO NOT switch the output polarity of the calibrator.
Perform a –10V DC gain calibration. Press
2
calibration menu. Press
procedure. Be sure to allow time for thermal offsets to stabilize
(usually about 1 minute).
Use the knob
and
3
difference of the calibrator output and the measured offset (from step 3)
and press
. Using the previous example values, enter 10 µV
minus 10.001 volts or –10.000990 volts.
When the adjustment finishes, verify that new readings fall
4
within ± 30 µV of the calibrator output minus the offset.
86
again to begin the adjustment
to set the number in the display to the