Agilent Technologies 34970A Service Manual page 29

Data acquisition/switch unit
Hide thumbs Also See for 34970A:
Table of Contents

Advertisement

Chapter 1 Specifications
To Calculate Total Measurement Error
Understanding the " % of range " Error The range error compensates
for inaccuracies that result from the function and range you select.
The range error contributes a constant error, expressed as a percent of
range, independent of the input signal level. The following table shows
the range error applied to the DMM 's 24-hour dc voltage specification.
Range
Input Level
10 Vdc
10 Vdc
10 Vdc
1 Vdc
10 Vdc
0.1 Vdc
Total Measurement Error To compute the total measurement error,
add the reading error and range error. You can then convert the total
measurement error to a "percent of input" error or a "ppm (part-per-
million) of input" error as shown below.
% of input error
=
ppm of input error =
Example: Computing Total Measurement Error
Assume that a 5 Vdc signal is input to the DMM on the 10 Vdc range.
Compute the total measurement error using the 90-day accuracy
specification of
(0.0020% of reading + 0.0005% of range).
±
Reading Error
= 0.0020% x 5 Vdc
Range Error
= 0.0005% x 10 Vdc = 50
Total Error
= 100
Range Error
Range
(% of range)
Error Voltage
0.0004
0.0004
0.0004
Total Measurement Error
Input Signal Level
Total Measurement Error
Input Signal Level
= 100
V + 50
V
=
µ
µ
±
=
±
=
±
40
V
µ
40
V
µ
40
V
µ
× 100
× 1,000,000
V
µ
V
µ
150
V
µ
0.0030% of 5 Vdc
30 ppm of 5 Vdc