Quickie Puma 40 Instructions For Use Manual page 32

Hide thumbs Also See for Puma 40:
Table of Contents

Advertisement

Available languages

Available languages

FAULT CODE
FLASH POSSIBLE CAUSE
1
2
3
4
5
6
7
8
9
10
10 + S
10 + A
32
Puma 40 Sedeo Ergo
The batteries need charging, or there is a bad connection to the battery. Check the
connections to the battery. If the connections are good, try charging the batteries.
The left hand motor has a bad connection. Check the connections to the left hand motor.
The left hand motor has a short circuit to a battery connection. Contact your Sunrise Medical
authorised dealer.
The right hand motor has a bad connection. Check the connections to the right hand motor.
The right hand motor has a short circuit to a battery connection. Contact your Sunrise Medical
authorised dealer.
The wheelchair is being prevented from driving by an external signal. One possibility is that the
battery charger is plugged in.
A joystick fault is indicated. Make sure that the joystick is in the centre position before
switching on the control system.
A control system fault is indicated. Make sure the control system connections are secure.
The parking brakes have a bad connection. Check the parking brake and the motor
connections. Make sure that the control system connections are secure.
An excessive voltage has been applied to the control system. This is usually caused by a poor
battery connection. Check the battery connections
S = Speed indicator LED's
A communication fault is indicated. Make sure that the joystick cable is securely
connected and not damaged.
A = Actuator LED's
An actuator trip is indicated. If more than one actuator is fitted, check which actuator
is not working. Check the actuator wiring.
Fig.8.2
Rev.1.0

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sedeo ergo

Table of Contents