Error Codes; Determine The Problem - WEIGHSOUTH WSI-602 Instruction Manual

Table of Contents

Advertisement

Cleaning
• Disconnect the power before cleaning.
• Use a cloth with mild suds and light cleaning agents.
• Make sure that fluid is not able to get into the scale.
• Use a clean and soft cloth for rub off.

10.2. Error Codes

Error Code Description
Err 4
Err 6

10.3. Determine the Problem

Determine whether the problem is in the PCB or the Load Cell
• Remove power from the system, and disconnect the load cell
connection from the PCB
• Connect the PCB to a load cell simulator
• Reapply power and test the PCB
• If problem goes away, its source is probably in the Load cell.
Check the wiring, connecter, load cell and mechanical com-
ponents of the load cell.
If problem persists, its source is probably in the PCB. Check the
PCB voltages, connecters, cables and function programs
Zero range exceeded.
A/D Count out of the
range
POSSIBLE CAUSES
• Goods on the platform
• Overload, when zeroing the
scale.
• Improper calibration
• Load cell problem
• PCB problem
• Platform not installed
• Load cell problem
• PCB problem