Error Codes

1. Overload
2. Wrong connection with Load Cell
UUUUUU
3. Load cell has quality issue.
1. Calibration is off
nnnnnnn
2. Wrong connection
3. Load cell has quality issue
ERR1
During calibration, not input the weights or
the weight is overload
ERR2
During calibration, the input signal is nega-
tive
ERR3
During calibration, the input signal is nega-
tive
ERR4
During calibration, the signal is unstable
Visit www.balances.com to purchase your WeighSouth VS2501
1. Reduce Weight
2. Confi rm that the Base is Con-
nected to the Indicator. Check
for any cuts in the cable
Replace Load Cell
1. Check scale feet to see if
they are level and Base is set-
ting fi rm on surface.
2. Check load cell connection.
3. Checking load cell check
input and output resistance
1. Input the correct weights
The calibration weights Mini-
mum is 10% of Max. cap.
Recommend 60%-80% of Max.
Cap.
1. Check the connection of the
Base.
2. Confi rm that the corect cali-
bration data was inputted into
the Indicator.
After the platform is stable, start
calibration