Operating Environment • 45 to 95° F (7 to 35° C)
Table 1. Models 1028 and 1029 Technical Specifications

Specifications

Windows 8.1, Windows 7, or Windows XP (32-bit or 64-bit)
Note: Windows XP is not recommended since Microsoft no lon-
ger supports it.
Minimum
• Processor: x486
• Total RAM: 128 MB
• Disk space: 10 MB
• USB port: one
• Display resolution: 1024 x 768
• Color depth: 32-bit
0.1 to 9999 picocuries/liter (pCi/l) or 1 becquerels per cubic meter
3
(Bq/m
) to 99.99 kilo becquerels per cubic meter (kBq/m
±25% or 1 pCi/l, whichever is greater after 24 hours
Diffused-junction photodiode
• Model 1028—quantity 1
• Model 1029—quantity 2
Active volume—9.4 cm
Dome volume—63 cm
0.5 (1029 only), 1, 2, 4, 8, 12, 16, 20 or 24-hour intervals, select-
able by user
1, 12, 24, 36, 48, 60, 72, 84, 96, 100, 999 hours, selectable by
user (maximum of 720 measurements)
If 999 is selected, measures until memory is full or to 720 data
points, whichever occurs first
• 1028—3 counts per hour per picocurie per liter (cph/pCi/l)
• 1029—6 counts per hour per picocurie per liter (cph/pCi/l)
16-digit reflective LCD display
Green LED (light emitting diode) indicates radon detections
6-digit membrane switches with multiple functions
Inertial switch
• Temperature—0 to 120 ± 2 degrees F
• Pressure—10 to 15 ± 0.5 psi
• Humidity—20 to 90% relative humidity ± 5%
• 20 to 80% relative humidity, non-condensing