Remove the test leads from the calibrator
before you open the battery door.
Make sure the battery door is closed and
latched before you operate the calibrator.
Do not operate the calibrator if it is damaged.
Do not operate the calibrator around explosive
gas, vapor, or dust.
To avoid a violent release of pressure in a
pressurized system, shut off the valve and
slowly bleed off the pressure before you attach
or detach the calibrator pressure fitting to the
pressure line.
When servicing the calibrator, use only
specified replacement parts
To avoid overpressure damage, do not apply
pressure that exceeds the following:
Model 713 30G: 3X top of range (90 psi, 620 kPa,
6.2 bar)
Model 713 100G: 2X top of range (200 psi, 1380
kPa, 13.8 bar)
To avoid corrosion in the pressure sensor, use the
calibrator only with media compatible with glass,
ceramic, silicon, RTV, type 316 stainless steel, and
nickel.
To avoid mechanically damaging the calibrator, do
not apply torque between the pressure fitting and
the calibrator case. See the figure for the proper
use of tools.
Caution