Keithley 2700 User Manual page 66

Multimeter/switch system
Hide thumbs Also See for 2700:
Table of Contents

Advertisement

Model 2700 Multimeter/Switch System User's Manual
Figure 1-10
Exercise 6 — Return a single reading (continuous triggering)
Return Readings
1. If a MATH function (mX+B, percent or 1/X) is enabled, the result of the calculation will
be returned. If there is no MATH function enabled, FETCh? and CALC:DATA? will return
the basic reading. MATH functions are covered in Section 5.
2. None of these read commands trigger measurements. They simply return the lastest
reading. If FETCH?, CALC:DATA? or DATA? is again sent before a new reading is
triggered, the old reading will be returned.
3. DATA:FRESh? can only be used once to return the same reading. Sending it again before
a new reading is triggered will cause error -230 (data corrupt or stale).
Trigger Configuration
FETCh?
CALC:DATA?
Return result of MATH
calculation
SAMP:COUN 1
INIT:CONT ON
DATA?
OR
DATA:FRESh?
Return Basic Reading
1, 2
Getting Started
Place 2700 in
continuous
trigger state.
OR
2, 3
1-37