15-14
SCPI Reference Tables
Table 15-5 (continued)
SENSe command summary
Command
[SENSe[1]]
:CURRent:AC
:APERture <NRf>
[, <clist>]
:APERture? [<clist>]
:NPLCycles <n>[, <clist>]
:NPLCycles? [<clist>]
:RANGe
[:UPPer] <n>[, <clist>]
[:UPPer]? [<clist>]
:AUTO <b> [, <clist>]
:AUTO? [<clist>]
:DIGits <n> [, <clist>]
:DIGits? [<clist>]
:REFerence <n>[, <clist>]
:STATe <b> [, <clist>]
:STATe? [<clist>]
:ACQuire [, <clist>]
:REFerence? [<clist>]
:AVERage
:TCONtrol <name>
:TCONtrol?
:WINDow <NRf>
:WINDow?
:COUNt <n> [, <clist>]
:COUNt? [<clist>]
[:STATe] <b> [, <clist>]
[:STATe]? [<clist>]
:DETector
:BANDwidth <NRf>
[, <clist>]
:BANDwidth? [<clist>]
Model 2700 Multimeter/Switch System User's Manual
Description
Path to configure AC current.
Set integration rate in seconds (60Hz;
1.67e-4 to 1, 50Hz; 2e-4 to 1).
Query aperture integration rate.
Set integration rate in line cycles (60Hz;
0.01 to 60, 50Hz; 0.01 to 50).
Query line cycle integration rate.
Path to set measurement range:
Select range (0 to 3.1).
Query range.
Enable or disable auto range.
Query state of auto range.
Specify measurement resolution (4 to 7).
Query resolution.
Specify reference (-3.1 to 3.1).
Enable or disable reference.
Query state of reference.
Use input signal as reference.
Query reference value.
Path to configure and control filter:
Select filter type: (MOVing or REPeat).
Query filter type.
Set filter window in % of range
(0 to 10).
Query filter window.
Specify filter count (1 to 100).
Query filter count.
Enable or disable filter.
Query state of digital filter.
Path to set bandwidth:
Set AC detector bandwidth in Hertz
(3 to 3e5).
Query bandwidth.
Default
parameter
Ref
SCPI
Sec 3
(Note 2)
Sec 4
5.0
Sec 4
Sec 4
3
ON
6
Sec 4
0
Sec 5
OFF
Sec 4
(Note 3)
0.1
10
(Note 4)
Sec 4
30