SunTech 247 BP+VITALS User Manual

SunTech 247 BP+VITALS User Manual

Blood pressure device for automated and manual measurement
Hide thumbs Also See for 247 BP+VITALS:
Table of Contents

Advertisement

Quick Links

Us e r Guid e

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 247 BP+VITALS and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for SunTech 247 BP+VITALS

  • Page 1 Us e r Guid e...
  • Page 2 SunTech 247 ® Blood Pressure Device for Automated and Manual Measurement User‘s Manual...
  • Page 3: About This Manual

    About this Manual ® This manual describes features and uses of the SunTech 247 by SunTech Medical , Inc., a non-invasive, clinical-grade automated device to measure blood pressure, pulse rate and mean arterial pressure. Optional modules to measure temperature and functional oxygen saturation are available, and their use is also described in this manual.
  • Page 4: Changes And Reissues

    Changes and Reissues This manual is identified as Part Number: 80-0040-00. Changes occurring between issues of this document are addressed through change information sheets, addenda, or replacement pages. If none of these accompany this manual, the manual is correct as printed. Should you notice errors or omissions in this manual, please notify us at: SunTech Medical 507 Airport Boulevard, # 117...
  • Page 5: Keys And Acronyms

    KEYS AND ACRONYMS Acronyms Acronyms commonly used in this document include: SunTech proprietary acronym for ―All Purpose Cuff‖ Blood pressure Pulse rate K-sound Korotkoff sound Mean arterial pressure NIBP Non-invasive blood pressure EMR/EHR Electronic Medical Records/Electronic Health Records Document Key This manual uses the following icons to call attention to specific instructions or guidance.
  • Page 6: Indications For Use

    Indications for Use SunTech 247 NIBP , Temperature, and Pulse Oximeter device is indicated for use in measuring and displaying Systolic and Diastolic blood pressures, pulse rate, temperature, and functional oxygen saturation (SpO ) of adult and pediatric patients in hospitals, medical facilities, clinics, physicians offices, and other sub-acute environments.
  • Page 7 For accurate blood pressure measurements, ensure that the circumference of the arm fits within the range markings on the cuff. SunTech 247 is not intended for continuous monitoring. Although the blood pressure cuff and cable are defibrillator proof, the temperature probe and SpO sensor are not.
  • Page 8 SunTech Do not make repairs yourself. Equipment must be returned to authorized service personnel for repairs. Substitution of a component different from that supplied may result in measurement error. SunTech 247 If the is dropped or mishandled, please have it checked by a authorized service center before bringing it back into use.
  • Page 9: Table Of Contents

    TABLE OF CONTENTS Changes and Reissues ....................iii Copyright Information ....................iii KEYS AND ACRONYMS ..................IV Acronyms ..........................iv Document Key ........................iv Indications for Use ......................v User Responsibility ......................v Warnings and Contraindications ..................v TABLE OF CONTENTS ................... VIII SUNTECH 247 GETTING TO KNOW THE ............
  • Page 10 Safety Precautions ......................16 Protecting Your Patient ....................16 Protecting Yourself ......................16 SunTech 247 Protecting Your ..................16 Mounting Your Device ...................... 17 Mounting the Device on a Wall ..................17 Affixing the Device to a Mobile Stand................17 Placing the Device on a Tabletop ...................
  • Page 11 Measuring Oxygen Saturation with Sensor on Finger ............30 Measuring Oxygen Saturation with a Earlobe Sensor ............31 MANAGING READINGS ...................32 Recalling the Last Set of Readings ..................32 Clearing the Last Set of Readings ..................32 SUNTECH 247 MAINTAINING THE ..............33 Routine Maintenance .....................
  • Page 12 APPENDICES....................50 Specifications ........................50 Compliance ........................51 Safety Requirements ...................... 51 EMC Statement ......................52 Symbols ..........................52 EU and FCC Regulatory Information ................58 Limited Warranty ......................59 Purchasing Parts and Accessories ..................61...
  • Page 13: Getting To Know The Suntech 247

    SUNTECH 247 GETTING TO KNOW THE SunTech 247 Your is a model of functionality, offering consistent blood pressure measurements along with reliable temperature and functional oxygen saturation readings. Package Contents SunTech 247 is available in two versions – with and without a rechargeable battery. Contents for each version are listed below.
  • Page 14: The Suntech 247 Battery Kit

    SunTech 247 Battery Kit SunTech 247 Battery kit consists of a rechargeable battery powered device. The SunTech 247 remainder of its contents matches the kit. Accessory Modules SunTech Accessory modules that you can purchase from your distributor or to enhance SunTech 247 usability of your include:...
  • Page 15: A Bird's Eye View

    A Bird’s Eye View BP module Automatic button Power button (recessed) SphygMode™ manual button Blood Power pressure hose connector connector (rear panel) Connectors on the main BP module  Power Connector: Connects to the power supply.  Blood Pressure Hose Connector: Connects to the 8-foot pressure hose. Buttons The buttons are used for all actions on the interface, and provide tactile feedback when pressed.
  • Page 16 Buttons on the main BP module Button Functionality for the BP Module Button Device Status Action Result Select for less than 2 Start an automatic BP seconds measurement. Select for 2 to 5 Redisplay last measurement values Idle seconds for all modules. Select for more than 5 Clear last measurement values for seconds...
  • Page 17: Bp Display

    BP display Systolic blood pressure Pressure column Diastolic blood pressure Mean arterial Pulse rate icon pressure icons Pulse rate/ Mean arterial pressure/Status code Warning icon Cuff/hose Power level error icon indicator Information on the BP module display  Systolic blood pressure: At the end of a measurement or when the last measurement is recalled, the patient‘s systolic BP is displayed.
  • Page 18: Optional Temperature And Pulse Oximetry Modules

     Warning icon: When displayed, indicates that the system needs to be checked. See page 40 for details.  Pressure column: Displays the pressure in the cuff. Each segment represents approximately 10mmHg. Optional temperature and pulse oximetry modules Temperature units select (recessed) Temperature mode select...
  • Page 19 Connectors on the optional temperature module  Temperature unit connector: Located on the temperature module, connects to the temperature probe housed in the temperature probe well. Temperature probe connector Temperature unit connector (not shown) Buttons on the optional pulse oximetry module: None Connectors on the optional pulse oximetry module ...
  • Page 20: Optional Temperature And Pulse Oximetry Displays

    Optional temperature and pulse oximetry displays Celsius icon Fahrenheit icon Temperature Warning icon Signal quality Warning icon Information on the optional temperature module display  Temperature: At the end of a measurement, the patient‘s temperature is displayed. During a measurement, displays the probe type and related information. If the warning icon is lit, displays the status code.
  • Page 21: Icons And Cues

    Information on the optional pulse oximetry module display  : At the end of a measurement, the functional oxygen saturation of the patient is displayed. If the warning icon is lit, displays the status code. See page 42 for details. ...
  • Page 22: Visual Cues - Battery Icon

    Visual Cues - Battery Icon SunTech 247 If your does not have a battery, all segments of the battery icon are lit when the power supply is connected. If your device has a battery, then the icon indicates the status of the power supply as follows.
  • Page 23: Visual Cues - Temperature Module

    ICON/DISPLAY INDICATES… If this icon is lit on power-up, MAP mode is enabled. After a measurement, this icon is lit when MAP is displayed in the space below. These letters are displayed in the pulse rate display area when you are checking the device‘s calibration.
  • Page 24: Visual Cues - Pulse Oximetry Module

    ICON/DISPLAY INDICATES… The device is set to measure an oral temperature. The device is set to measure an axillary temperature. The device is set to measure a rectal temperature. The unit is taking a measurement in direct measurement mode. The display of alternates with the current probe temperature.
  • Page 25 ICON/DISPLAY INDICATES… Indicates signal strength and quality from the pulse oximeter sensor. If there is no measurement and the signal quality is low, try a different site or sensor. Indicates the functional oxygen saturation. value This area also displays the status code when the warning symbol is lit.
  • Page 26: Quick Start Guide

    QUICK START GUIDE If the device is off, turn it on by depressing the power button on the right side. Measuring Blood Pressure and Pulse Rate Automatically Wrap an appropriately sized cuff (sizes are tabulated on page 21) snugly around the upper arm midway between the elbow and shoulder.
  • Page 27: Measuring Temperature

    Measuring Temperature To measure temperature orally, remove the blue probe from its holder and secure a disposable cover on it. The display shows the type of measurement, Wait for the device to beep before placing the probe carefully under the patient‘s tongue as denoted by the heat pockets shown to the right.
  • Page 28: Setting Up The Suntech 247

    SUNTECH 247 SETTING UP THE Safety Precautions SunTech 247 As a clinically trained professional using the , your responsibilities include safeguarding your patients, yourself and your equipment. Many setup functions will be performed either only once or very occasionally, and it is important that you pay close attention.
  • Page 29: Mounting Your Device

    Mounting Your Device For convenience, you may mount your unit on the wall or attach it to a mobile stand or a tabletop stand. A storage basket is included and can be used to hold cuffs, boxes of probe covers for the optional temperature module, and SpO sensors for the optional pulse oximetry SunTech 247 module.
  • Page 30: First-Time Setup

    First-Time Setup SunTech 247 must be plugged in and charged before first use. Charging the Battery SunTech 247 Battery is powered by a rechargeable 6V lead-acid battery or by AC power. To turn the device on for the first time, connect the device to the power supply, then the power supply to an AC mains power source.
  • Page 31: Selecting Temperature Unit Of Measurement

    For SpO measurements, attach the pulse oximetry sensor or extension cable to the connector on the module, making sure the retention clip is pushed towards the front of the unit. Once the sensor/cable connector is inserted, push the clip backward over the connector to hold it securely in the module.
  • Page 32: Measuring Blood Pressure With The Suntech 247

    SUNTECH 247 MEASURING BLOOD PRESSURE WITH THE SunTech 247 Your device is designed to take accurate blood pressure readings by the oscillometric method. Systolic pressures from 60 to 270 mmHg and diastolic pressures from 30 to 170 mm Hg lie within the range of your device. In most cases, you will be able to take accurate blood pressure (BP) and pulse rate (HR) measurements within 30-40 seconds.
  • Page 33: Selecting The Right Cuff

    Selecting the Right Cuff SunTech Medical, Your device comes with durable two-piece All Purpose Cuffs (APC) from Inc. Child Thigh SunTech 247 . Cuffs are available in a range of sizes, from . Note that your works optimally with APC cuffs. Using the table below, select a cuff you estimate to be of the right circumference: Wrap the cuff around the patient‘s upper arm midway between the elbow and the shoulder.
  • Page 34: Taking A Measurement

    Taking a Measurement SunTech 247 allows you to take BP measurements automatically like a monitor or SphygMode manually like a sphygmomanometer using manual In automatic mode, the cuff inflates and deflates automatically. Initial inflation reaches a cuff pressure of 160 mmHg; the cuff then re-inflates as necessary to obtain a reading. Deflation is optimized to reduce measurement time and obtain an accurate result.
  • Page 35: Taking A Sphygmode Measurement

    If there is an error in obtaining a measurement, indicated by three beeps, please refer to the troubleshooting tips on page 40 and take the appropriate remedial measure. Taking a measurement on each arm helps rule out dissecting aneurysms, coarctation of the aorta, vascular obstruction and possible errors in measurement.
  • Page 36: K-Sounds: A Primer

    K-Sounds: A Primer Korotkoff sounds, commonly called K-sounds, are the sounds you detect through your stethoscope when you measure blood pressure with a sphygmomanometer or an aneroid device. Named for the Russian physician who identified them, there are five phases of K- sounds, each phase characterized by a distinct volume and quality of sound.
  • Page 37: Measuring Temperature With The Suntech 247

    SUNTECH 247 MEASURING TEMPERATURE WITH THE SunTech 247 Your device can measure temperature with the optional temperature module. This module enables you to take rapid, accurate temperature measurements ranging from 86°F-109.4°F. Typically, predictive readings are obtained within ten to fifteen seconds, and direct readings within two minutes.
  • Page 38: Taking An Oral Temperature

     Use only FILAC™ FASTEMP™ probe covers with this device.  The device and probe covers are non-sterile. Do not use on abraded tissue.  To limit cross contamination, use blue probes for taking oral and axillary temperature only. Use red probes for rectal temperatures only. ...
  • Page 39: Taking An Axillary Temperature

    Taking an Axillary Temperature Remove the blue probe from its holder and secure a disposable cover on it. The probe pre-heating process begins with the flashing display of the type of measurement, Briefly (less than 2 seconds) press the temperature mode select button until it displays for axillary.
  • Page 40 Remove the red probe from its holder and secure a disposable cover on it. The probe pre-heating process begins with the flashing display of After a few seconds, the device will beep to signify that the probe is ready for placement.
  • Page 41: Measuring Oxygen Saturation With The Suntech 247

    SUNTECH MEASURING OXYGEN SATURATION WITH THE SunTech 247 optional pulse oximeter modules measure functional oxygen saturation ranging from 40% to 100%. A signal strength display assists the clinician in the proper placement of the sensor. SunTech 247 is a spot check device and is not used for patient monitoring.
  • Page 42: Guidelines For Use

     Do not use a damaged sensor as it may cause patient injury or equipment failure.  The use of this sensor is contraindicated in patients with allergies to adhesive tape. Guidelines for Use  When selecting a sensor site, give priority to an extremity free of an arterial catheter, blood pressure cuff or intravascular infusion line.
  • Page 43: Measuring Oxygen Saturation With A Earlobe Sensor

    When you remove the sensor from the patient‘s finger, the display will flash the last measurement for 8 seconds. The measurement will then be displayed for 2 minutes or until another measurement is made. Note the patient‘s reading and confirm normal venous return.
  • Page 44: Managing Readings

    MANAGING READINGS Recalling the Last Set of Readings To redisplay the last set of readings, depress the automatic button on the BP module for SunTech 247 more than two seconds until the last reading is displayed. If your temperature and/or pulse oximetry modules, the last set of readings includes these readings as well.
  • Page 45: Maintaining The Suntech 247

    SUNTECH 247 MAINTAINING THE Routine Maintenance SunTech Establishing simple care guidelines helps protect the performance and life of your . On a routine basis, you should inspect the device, cables and pneumatic hoses for cracks, fraying or kinks and immediately replace any damaged parts. Remember to check the calibration of the BP module every two years.
  • Page 46: Checking The Calibration Of Your Device

    Dry sensors and clips thoroughly before re-use. Never immerse sensors and clips in fluids. Do not pour or spray any liquids on them either. Caustic or abrasive cleaners will cause permanent damage. Prying the finger clip sensor to an angle greater than 90° can permanently damage its casing.
  • Page 47: Replacing The Rechargeable Battery

    1. Remove the Temperature Well from the Temperature Module. 2. Replace the Temperature Probe with the Calibration Plug. 3. Replace the Temperature Well. 4. Initiate a measurement by inserting and removing the Temperature probe from the well. 5. Verify the accuracy of the measurement. Probe accuracy can be checked by using the thermometer as you would on a patient in any mode with a cover, but place it in a calibrated water bath.
  • Page 48: Attaching The Temperature Module

    Attaching the Temperature Module Should you need to attach or remove the temperature module, the following instructions give an overview of its attachment to the BP module. The temperature module attaches to the top of the BP module. It is made up of the following two pieces: ...
  • Page 49 Attach the temperature probe connector to part B as follows: with the black surface facing downward, hold the probe connector end against the notches on the probe holder, as shown in red (step 1 shown below). Rotate the connector upwards until it snaps securely into place (step2 shown above). The black surface faces outward and the cord extends upward.
  • Page 50: Attaching The Pulse Oximetry Module

    Attaching the Pulse Oximetry Module Should you need to attach or remove a pulse oximetry module, the following instructions give an overview of its attachment to the BP module. The pulse oximetry module attaches to the bottom of the BP module. Using the power button on the right side of the BP module, ensure SunTech 247 that the...
  • Page 51: Storage, Shutdown, Transport

    Storage, Shutdown, Transport Storage SunTech 247 must be stored between -20C (-4F) and 55C (131F). Relative humidity must be less than 90%. SunTech 247 If you are storing the Battery for 30 days or longer, it is recommended that you disconnect the battery from the device.
  • Page 52: Troubleshooting

    TROUBLESHOOTING The troubleshooting chart provides pointers on diagnosing issues associated with error or status codes. Problem: The SunTech 247 will not power on. Solutions: SunTech 247 1. The must be plugged in to charge before first use. Problem: My SunTech 247 is not communicating with my EMR/EHR system or network Solutions: Bluetooth 1.
  • Page 53: Troubleshooting - Temperature Module

    Problem: Air leak, Loose cuff, or Blocked or pinched hose Solutions: 1. Check that the hose has no sharp bends or is pinched. 2. Check that the patient is not lying on the cuff. 3. Check that the cuff is in the correct position. 4.
  • Page 54: Troubleshooting - Pulse Oximetry Module

    When properly fixed, the error should no longer be displayed. Take a new measurement. Status Code: 10 Problem: Defective temperature probe Solutions: 1. Remove and then place the probe in the well. 2. Ensure that the temperature well (the rear portion of the temperature module) is well seated on the display.
  • Page 55 Solution: Remove the sensor from the patient. Redisplay the last measurement prior to the timeout, or take a new measurement by placing the sensor on the patient. Status Code: 02 Problem: Poor sensor position (signal is inadequate for a reliable measurement) Solution: Adjust position of sensor on patient by placing sensor on opposite hand or ear or alternate site.
  • Page 56: Faqs

    FAQs SunTech 247 Can I obtain replacement copies of the CD and manual? SunTech 247 Copies of the manual are available through the Customer Service area of our website. To download the manual, please visit http://www.SunTechMed.com and navigate to Customer Service. For a replacement CD, please email Customer Service at CustomerService@SunTechMed.com SunTech 247...
  • Page 57 supplements this with the ability to take measurements as you would if you were using a mechanical sphygmomanometer. SunTech 247 Can I upgrade my current version of at a later date? SunTech 247 To upgrade your device, please review the list of accessories on page 61 or on our website.
  • Page 58: Web Resources

    When does the warranty period begin? SunTech 247 The warranty for your begins on the date of shipment of your device. SunTech 247 How do I make a warranty claim for the Contact an authorized service center on pages 46 - 46. SunTech 247 How do I get my to communicate with my EMR/EHR software system?
  • Page 59: For Customers In Europe, The Middle East, And Africa

    For customers in Europe, the Middle East, and Africa SunTech Medical, Ltd. Oakfield Estate Eynsham, Oxfordshire England Tel: +44 (0) 1865.884.234 Fax: +44 (0) 1865.884.235 For customers in Asia and the Pacific SunTech Medical, Ltd. Level 19, Two, International Finance Centre 8, Finance Street, Central Hong Kong Tel: +852.2251.1949...
  • Page 60: Special Situations

    SPECIAL SITUATIONS Special Situations Unique circumstances, such as the patient‘s age or physiological disturbances, require you to take special care while measuring blood pressure or vital signs. The more common examples SunTech 247 of such circumstances are described here, to assist you in using your optimally under such conditions.
  • Page 61: Measuring Blood Pressure In The Elderly

    SunTech 247 SphygMode For clinically relevant hypertension in pregnancy, use the to take a measurement. Measuring Blood Pressure in the Elderly In the elderly, the combination of hypertension and ageing can manifest as a decrease in arterial compliance. Variability in blood pressure can lead to a number of circadian blood pressure patterns that are best identified using ambulatory blood pressure measurement.
  • Page 62: Appendices

    APPENDICES Specifications Patient population: Adult and pediatric patients (age 3 and above). Method of measurement: Oscillometric Initial inflation pressure: 160mmHg +/- 20mmHg Blood pressure range (mmHg): 60< Systolic BP< 270, 30< Diastolic BP< 170 Blood pressure accuracy: Measurements determined with this device are equivalent to those obtained by a trained observer using the cuff/stethoscope auscultation method, within Manual, electronic or the limits prescribed by the American National Standard,...
  • Page 63: Compliance

    Safety systems: Independent hardware over-pressure circuit and redundant software overpressure algorithm to limit cuff pressure to less than 330 mmHg. Independent hardware timing circuit and redundant software timer algorithm to limit the duration of a blood pressure cycle to less than 180 seconds. Dimensions: Length = 5.5 inches, Height = 11.5 inches, Width = 3.8 inches;...
  • Page 64: Emc Statement

    EMC Statement This equipment has been tested and found to comply with the limits for medical devices to IEC60601-1-2: 2001. These limits are designed to provide reasonable protection against harmful interference in a typical medical installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to other devices in the vicinity.
  • Page 65 SYMBOL DESCRIPTION No SpO alarm Attention, consult accompanying documents. Class II isolation equipment CE approval Recognized component certified by UL to both Canada and US requirements TUV Canada and US approval TUV International approval Earth ground Output connection configuration – positive voltage;...
  • Page 66 Guidance and manufacturer‘s declaration – electromagnetic emissions SunTech 247 is intended for use in the electromagnetic environment specified below. SunTech 247 The customer or the user of the should assure that it is used in such an environment. Emissions test Compliance Electromagnetic environment –...
  • Page 67 Guidance and manufacturer‘s declaration – electromagnetic immunity SunTech 247 is intended for use in the electromagnetic environment specified below. SunTech 247 The customer or the user of the should assure that it is used in such an environment. Immunity test IEC 60601 test Compliance level Electromagnetic environment –...
  • Page 68 Guidance and manufacturer‘s declaration – electromagnetic immunity SunTech 247 device is intended for use in the electromagnetic environment specified SunTech 247 below. The customer or the user of the device should assure that it is used in such an environment. Immunity test IEC 60601 Compliance...
  • Page 69 Recommended separation distances between portable and mobile RF communications SunTech 247 equipment and the device SunTech 247 device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the SunTech 247 device can help prevent electromagnetic interference by maintaining a minimum distance between...
  • Page 70: Eu And Fcc Regulatory Information

    Bluetooth For units that are ® enabled, the following EU and FCC regulatory information applies: EU Regulatory Information The transmitter module manufactured by Mitsumi and incorporated into the 247 BP module marked by has been approved in accordance with the R&TTE directive. FCC Regulatory Information FCC RF Interference Statement: This equipment has been tested and found to comply with the limits pursuant to Part 15 of...
  • Page 71: Limited Warranty

    Limited Warranty SunTech 247 Device SunTech Medical, Inc. provides the original purchaser the following limited warranty from date of invoice. All serialized modules 24 months APC Cuff(s) 24 months Roll stand 24 months Accessories, i.e. patient cables, disposables, power 90 days supplies SunTech Medical, Inc.
  • Page 72 The instrument will be repaired in the shortest possible time and returned prepaid by the same shipping method as received by the factory. This limited warranty is void if the instrument has been damaged by accident, misuse, negligence, act of God or serviced by any person not authorized by SunTech Medical, Inc.
  • Page 73: Purchasing Parts And Accessories

    Purchasing Parts and Accessories SunTech 247 We recommend that you purchase parts and accessories for your from your authorized SunTech 247 distributor. A consolidated list of parts and accessories appears below. All Purpose - General Clinical Use Cuffs Item # Item Name Item Description 98-0223-00...
  • Page 74 98-0080-08 All Purpose Cuff, Large Blood Pressure Cuff with quick- Adult connect connector, Range: 31-40cm Color: Burgundy 98-0080-09 All Purpose Cuff, Large Blood Pressure Cuff with quick- Adult Long connect connector, Range: 31-40cm Color: Burgundy 98-0080-10 All Purpose Cuff, Thigh Blood Pressure Cuff with quick- connect connector, Range: 38-50cm Color: Brown...
  • Page 75 Thermometry Accessories (Kendall FASTEMP®) Item # Item Name Item Description 98-0155-00 Red Rectal Temperature Kit Includes rectal probe, well, and box of probe covers 52-0004-00 Blue Oral/Axillary Probe Oral/axillary probe with 9 ft (2.7 m) extended cord 98-0146-00 Blue Oral/Axillary Well Rear piece of the temperature module that holds the oral/axillary probe and box of probe covers...
  • Page 76 98-0149-00 Tabletop stand kit Includes power supply holder 98-0148-00 Wall mount kit Includes wall mountable basket 98-0150-00 Basket Wall mountable 19-0013-00 Power supply for the Input: 100-240 V, 50-60 Hz; Output: SunTech 247 +6 V; medical grade 19-0014-00 Power supply for the Input: 100-240 V, 50-60 Hz;...
  • Page 77 www.SunTechMed.com SunTech Medical, Inc. 507 Airport Boulevard, Suite 117 Morrisville, NC 27560 Tel: +1. 919.654.2300 +1. 800.421.8626 Fax: +1. 919.654.2301 SunTech Medical, Ltd. Level 19, Two, International Finance Centre 8, Finance Street, Central Hong Kong Tel: +852.2251.1949 Fax: +852.2251.1950 SunTech Medical, Ltd. Service Department Oakfield Industrial Estate Eynsham, Oxfordshire OX29 4TS...
  • Page 78 Nellcor® is a registered trademark of Nellcor Puritan Bennett LLC FILAC™ and FASTEMP™ are registered trademarks of Tyco Healthcare Group LP Bluetooth ® word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by SunTech Medical, Inc. is under license.