TESTO 625 Instruction Manual page 11

Humidity/temperature measuring instrument
Hide thumbs Also See for 625:
Table of Contents

Advertisement

Select the desired option with
1
:
· YES: Switches automatic probe detection on
(recommended).
· no: Switches automatic probe detection off.
no was selected:
Use
/
2
confirm with
Continue with objective
YES was selected:
- Automatic probe detection is started.
- Once a radio probe is found, the probe ID is
displayed. If no probe is found, NONE lights up.
If more than one radio probe is switched on:
If necessary, select the desired probe with
Possible causes why probes are not found:
· The radio probe is not switched on or the
battery of the radio probe is spent.
· The radio probe is outside the range of the
measuring instrument.
· Sources of interference are influencing the radio
transmission (e.g. reinforced concrete, metal
objects, walls or other barriers between
transmitter and receiver, other transmitters of the
same frequency, strong electromagnetic fields).
If necessary, rectify the possible causes for the
disruption to the radio transmission and start
configuration mode again.
Press
2
to set the probe ID manually and
.
T
O SET
to change to the next function.
5. Operation 29
and confirm with
.
A
O
UTO
FF
/
.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents