Care and Maintenance
60
Operating Instructions Moisture Analyzer MA37
Contaminated Equipment:
Health risk posed by product contamination due to product deposits and
collection of residue with microbial contamination .
Health risk from biological or microbiological substances .
Observe cleaning specifications .
Examine the cleaning results closely .
Cleaning the Heating Module
Danger of Burns:
The inner part of the heating module and parts of the sample chamber may be
extremely hot .
Wait until the heating module has cooled completely .
Avoid touching the heating element . To remove spray residues or deposits from the
heating element, use a weak solvent such as ethanol .
Unlocking the Heating Module
1) Open the heating module all the way .
Removing the Heating Module
2) Pull the heating module back up out of the guide .