Download Print this page

KERN CH50K500 Operating Instructions page 4

Hanging balance

Advertisement

7
Operation Instruction
7.1
Weighing
• The balance is switched on by pressing the "ON/OFF/TARE"
key.
• The balance display shows "0.00".
• Hang on the weighing object carefully.
• If the weighing object is heavier than the weighing range the
display indicates "E" (=Overload). Be sure to prevent over-
loading (chap. 2.2 "Inappropriate use").
• The balance is switched off by pressing the "ON/OFF/TARE"
key once.
7.2
Weighing units switch-over
Using the "UNIT" key you can switch over the display between
N, lb and kg. The
marks the active setting.
7.3
Taring
• Switch the balance on with " ON/OFF/TARE " key and wait
for "0" to appear in the display. Hang on the tare load care-
fully and press the " ON/OFF/TARE " key again.
• The balance display becomes "0.00". The weight of the tare
load has however been memorised internally. The item to be
weighed can now be placed carefully in the tare container
and the weight can be read off.
7.4 HOLD-key / animal weighing / peak value weighing
The HOLD-key may be used to activate various functions.
To select and activate these functions, please follow the se-
quence of operations below:
Switch on the balance and keep pressing the HOLD-key
until "H x" appears on the display.
The currently activated function is shown above "H1-H6"
and by a triangle on the display.
To select one of the functions, please press the
ON/OFF/TARE -key. The setting is stored by briefly press-
ing the HOLD-key.
7.4.1
Functions
HOLD function:
The HOLD function allows the fixed display of weighing values
for longer periods of time.
Depending on the operating range the ON/OFF switch, criteria
may be selected from 4 options (H1-H4).
Selecting the setting "H1" freezes the weighing values for
5 seconds after the HOLD-key was operated.
Selecting "H2" freezes the weighing values from operating
the HOLD-key until another key is pressed.
Selecting "H3" freezes the weighing values automatically
for 5 seconds.
Selecting "H4" freezes the weighing values automatically
until a key is pressed.
HOLD function is set in the following way:
Switch on the balance by pressing ON/OFF/TARE key.
Press and hold the HOLD key until the display shows the
required setting (check above „H1–H4")
Save the setting by brief pressing the HOLD key
Animal weighing function:
CH-BA-e-0927
This function is suitable for unsteady weighing procedures. A
mean value is derived from 16 weighing values.
The animal weighing function is set in the following way:
Switch on the balance by pressing the ON/OFF/TARE key.
Press and hold the HOLD key until „H x" is shown on the
display.
Press the ON/OFF/TARE key repeatedly until „H5" is
shown on the dispay.
Save the setiing by briefly pressing the HOLD key.
Once again press HOLD key.
On the left site, on top of the display, a triangle will start to
blink.
Hang on the weighing object.
During the triangle blinking, the weighing values are
loaded.
Then an average weight value is shown consisting of the
loaded values.
Peak value function:
This function displays the maximum load value (peak value) of
a weighing process.
Measuring frequency: 200ms
Attention:
Under no circumstances may the peak value exceed the
stated maximum load of the balance (risk of fracture!!).
How to set the peak value.
Switch on the balance by pressing the ON/OFF/TARE key.
Press and hold the HOLD key until „H x" is shown on the
display.
Press the ON/OFF/TARE repeatedly until „H6" setting is
shown on the display.
Save the setting by briefly pressing the HOLD key.
Hang on the weighing object.
The operation of the peak value function is shown by the
triangle on top left side of the display.
-
The negative value (
) is shown right after unloading the bal-
ance.
9

Advertisement

loading