SECTION III
Model 4145A
Table 3-2. Error Messages (Sheet 2 of 2)
Error Message
Meaning
Corrective
Action
Syntax error
An illegal name was entered
in
a GET or SAVE command.
The first character
of a file name must be
alphabetic,
and all succeeding
characters
must
be alphanumeric.
PLOT area is not specified,
contains one or more alpha-
betic characters,
or is miss-
ing coordinate
delimiters
(comma or space).
Re-enter
the
PLOT area correctly.
Busy
No data
Illegal file type.
Auto calibration
is being per-
formed.
PRINT
Key was pressed with no
measurement
data in the data
buffer.
Use P, D, or S for the file type.
Auto calibration
is performed
every 5
minutes.
Perform
a measurement.
Printer/
Plotter is
not
connected
Not
compatible
No printer
or plotter
is con-
netted
to the instrument,
or
the printer
or plotter
is not
set to LISTEN.
The discs used in the copy
operation
havedif
f erent sys-
tem labels, or one of the
discs is not a 4145A useable
disc.
Connect the printer
or plotter.
Copy cannot be performed.
Close the
The fixture
lid is open at the
Close the lid or change the setup.
Make
fixture
lid
start of a measurement
in which
_ the output voltage
may exceed
sure the system cable is properly con-
+42V or lid is open during the
netted.
User Mode (See Fig. 39).
output
The fixture
lid was opened
Close the fixture
lid and make the measure-
disabled,
during a measurement
in which
ment again.
Do not open the fixture
lid
close the
the output voltage
may exceed
during measurement.
Make sure the
fixture
lid
+42V.
system cable is properly
connected.
Emergency
Output was shut down to pre-
Make sure the setup and all connections
vent SMU damage.
are correct.
Recovered
Indicates
that there was a mo-
When a power loss occurs during measure-
from power
mentary
power loss.
ment,
output
is shut
down.
Press MEA-
down
SUREMENT
key to continue
the mea-
surement.
Step overflow
NO OF STEPS for VARl
Change START value,
STOP value, or
exceeds 512.
STEP value.
Disabled
A disabled softkey
was
The softkey cannot be used in the
function
pressed.
existing measurement
setup.
3-14