HP 4145A Operation And Service Manual page 44

Semiconductor parameter analyzer
Hide thumbs Also See for 4145A:
Table of Contents

Advertisement

Model 4145A
SECTION III
0 5
The
CRT
is sectioned
into five areas, as shown in
the figure.
Description
for each area is as follows :
0
1
0
2
0
3
0
4
0
5
Page Display Area :
The instrument's
various pages are displayed
in
this area.
When the PLOT key is pressed, only
this area is output
to the plotter.
Command Display Area :
The SAVE, GET, PLOT,
PRINT,
PURGE and
REPACK commands
are displayed in this area.
Keyboard Input Line :
All keyboard
(front
panel) entries
are displayed
in this
area.
Up to
60 characters
can
be
entered
but only 27 can be displayed.
To edit
the
displayed
characters,
use
the
BACK,
FORWARD,
DELETE and INSERT
keys.
When
the RECALL
key is pressed, the previous
entry
is re-displayed.
System Message Line :
Displays instructions,
error messages, and error
codes.
When
the
instrument
is turned
on,
instrument
status is displayed.
Softkey Prompt
Display Area :
Displays Softkey Prompts
(SKP).
Note
@ , 0,
and @ are erased by pressing
the CLEAR key.
Figure 3-5. Useable Display Area.
CRT
intensity
and focus
adjustment.
(Requires
a
sm.all flat-tip
screwdriver.)
(1)
Turn on the 4145A.
The MENU
page will be
displayed on the CRT.
(2)
Press the EXTN softkey
and the DIAG softkey.
(3)
Press the GDU TEST softkey.
Display
will .be
as shown in the figure.
(4)
Adjust
INTENSITY
@
until
all three
lines
(each line has a different
intensity)
are visible.
(0)
(5)
Adjust
FOCUS
@
until
the corners
of the
displayed trace are sharp.
(6)
Press the MENU key.
Figure 3-6. Operator
Adjustment.
3-l 1
_--
-
_.
y-------x
.
--
---
-7
- -