3-10. Removing the LCD Module

After you have removed the case bottom and pca, use the following procedure to remove
the LCD module:
1. Remove the four Phillips screws from the corners of the LCD module.
2. Lift the LCD module out of the case top.

3-11. Removing the Elastomeric Keypad

After you have removed the case bottom and pca, lift the Elastomeric keypad out of the
case top.

3-12. Reassembling the Tester

Generally, reassembling the Tester is the reverse of disassembly. However, you must
follow special precautions when installing the pca and input module.

3-13. Installing the PCA and Input Module

1. With the pca held outside the case top and the flex cable connector latch pulled out,
3-6
1
Figure 3-2. Flex Cable Connector
insert the free end of the flex cable into the connector. Make sure that the flex cable
is fully engaged. Push the connector latch back against the connector body.
2
t4f.eps