USE
This paragraph describes the fundamental operating
procedures. Detailed information and operating procedures
that involve optional connections or options are supplied in
the paragraph "USE" of the appendix "Technical
Informations". Some procedures can be executed with
controller only when the accessory "Hand held terminal" is
available. Make all vacuum manifold and electrical
connections and refer to Turbo-V pump instruction manual
prior to operating the Turbo-V controller.
!
To avoid injury to personnel and damage to the equipment,
if the pump is laying on a table make sure it is steady.
Never operate the Turbo-V pump if the pump inlet is not
connected to the system or blanked off.
The following figure shows the hand held terminal
control panel.
1. Keyboard
push-button
selection. It is active only when the front panel
operation has been selected. Pressed once, the pump
runs at about 2/3 of the nominal speed. To unselect
the mode, press the push-button again.
2. Keyboard push-button for START, STOP, RESET
mode selection. It is active only when the front panel
operation has been selected. By pressing once the
starting phase begins; if pressed again it stops the
pump. If the pump has been stopped automatically by
a fault, this push-button must be pressed once to reset
the controller and a second time to restart the pump.
3. Keyboard push-button to recall on the display the
cycle number, cycle time and pumplife.
4. Keyboard push-button to recall on the display the
pump current, pump temperature, pump power and
rotational speed. It is always active regardless of the
operating mode selected. Push-buttons 3 and 4, if
pressed together for
controller in a routine where it is possible to program
some operation parameters.
5. LCD back-lighted alphanumeric display: dot matrix 2
lines x 16 characters.
Hand Held Terminal Control Panel
WARNING!
for
LOW
SPEED
mode
at least 2 seconds put the
USE PROCEDURE
Controller Startup
To startup the controller plug the power cable into a
suitable power source.
Starting the pump
To start the pump send the START command to related
pins of the input connector (pin 1 - 6 of P1 connector), or,
when the hand held terminal is available, press the START
pushbutton.
Pump Shutdown
To shutdown the pump remove the START command from
the related pins of the input connector (pin 1 - 6 of P1
connector), or, when the hand held terminal is available,
press the STOP pushbutton.
MAINTENANCE
The Turbo-V 150 PCB series controller does not require
any maintenance. Any work performed on the controller
must be carried out by authorized personnel.
When a fault has occurred it is possible to use the Varian
repair service. Replacement controllers are available on an
advance exchange basis through Varian.
Before carrying out any work on the controller,
disconnect it from the supply.
If a controller is to be scrapped, it must be disposed of
in accordance with the specific national standards.