Ohms Calibration - Fluke 5700A Series Service Manual

Multi-function calibrator
Hide thumbs Also See for 5700A Series:
Table of Contents

Advertisement

2-177. Two-Wire Compensation Power Supplies
The floating supplies for U7 and U8 consist of a dc-to-dc converter made up of U6, T1
and associated components. Switching-regulator control IC U6 switches +17S through
the primary of T1 at about 30 kHz as determined by R27 and C21. The switching creates
high-frequency spikes on the +17S line which are filtered out by L2, R25, C18, C19,
C20, and C52. When the two-wire compensation circuit is not in use, control line PB7
turns Q5 and Q6 off, which removes +17S, switching off U6. The secondary of T1 is
rectified and filtered by CR1, CR2, CR9, CR10, C22, and C23. The voltage is then
regulated by 8.2V zener diodes VR1 and VR2. Additional filtering is provided by L3, L4,
R43, R44, C24, and C25. Two capacitors (C53 and C54) between A COM and S COM
reduce switching noise that would otherwise appear across the calibration resistor.
2-178. High Side Cancellation
The high side voltage drop is eliminated by U7 and U8. Chopper-stabilized op amp U7
and buffer amplifier U8 supply up to 10 mA. The current from the high side of the
ohmmeter being calibrated enters at the OUTPUT HI binding post, goes through K1 and
K6 on the Ohms assembly, and is routed to the Ohms Cal assembly where it connects to
A COM. The A COM connection is a common for the floating supply that powers U7
and U8. The current then flows out of U8 through 500Ω in Z2 and through K29 to the
calibration resistor via the OHMS OUT HI line.
K29 connects the -input of U7 to the sense point of the calibration resistance via OHMS
SENSE HI. Relays K2 and K4 on the Ohms Main assembly connect the +input of U7,
AMP IN, to INT SENSE HI. Relays on the motherboard connect INT SENSE HI to the
OUTPUT HI or SENSE HI binding post. Connected this way, U7 controls the voltage at
the output of U8 so that the voltage at the input of U7 stays zero. This forces the voltage
drop in the path to zero. Diodes CR3-CR6 and resistor R45 provide protection for U7.
2-179. Low Side Cancellation
The voltage drop in the low side path is canceled by current from high current op amp
U9. U9 is driven by dual FET input op amp U10. The two non-inverting inputs of U10
are connected across the 500Ω resistor in Z2. This enables U10, in conjunction with the
four 10 kΩ resistors in Z2, to sense the current in the high side path and to supply
(through U9 and the 1 kΩ resistor in Z2) a current equal but opposite to the current
through the low side path. This cancels the drop in that path. This canceling current goes
through relay K28A to the low sense point of the calibration resistor via OHMS SENSE
LO. From there it goes through the output low path to the OUTPUT LO binding post and
then through the sense low path back to the Ohms Cal assembly and through K28B to S
COM.

2-180. Ohms Calibration

The remaining circuitry is used only for calibration of the resistor values. All values
except 1Ω and 1.9Ω are calibrated using a single external 10 kΩ resistance standard
connected to the binding posts. The 1Ω and 1.9Ω values are calibrated using an external
1Ω standard. How to do the procedure is described in Section 7 of the Operator Manual.
Theory of Operation
Analog Section Detailed Circuit Description
2
2-115

Advertisement

Table of Contents
loading

This manual is also suitable for:

5720a series

Table of Contents