cmline7=3
cmline8=Ge
cmline9=test
range=2048
rtpreset=1000
rtprena=0
ltpreset=1000
ltprena=0
roipreset=0
roiprena=0
roimin=0
roimax=2048
autoinc=0
datname=C:\MCA\DATA\SD0002.mcd
savedata=1
fmt=dat
caluse=1
calch00=1172.00
calvl00=1173.264000
calch01=1331.00
calvl01=1332.500000
caloff=-0.506315
calfact=1.000750
calfact2=0
calfact3=0
calunit=keV
roi=1156 1199
peak=1173.230000
roi=1323 1353
peak=1332.480000
The following commands perform actions and therefore usually are not included in a
WEBMCAx.CFG or .MCD file:
start
halt
cont
savecnf
ComTec GmbH
Windows Server Program
; subset of parameters as in a WEBMCAA.CFG file...
; Use Energy calibration
; Calibration points..
; Calibration formula polynomial coefficients..
; Calibration unit
; ROI limits..
; calibration peak in this ROI
:
; Clears the data and starts a new acquisition for system 1. Further
; execution of the .CTL file is suspended until any acquisition stops
; due to a preset.
; Stops acquisition of system 1 if one is running.
; Continues acquisition of system 1. If a time preset is already
; reached, the time preset is prolongated by the value which was valid
; when the „start" command was executed. Further execution of the ;.CTL
file is suspended (see start).
; Writes the settings into MCAA.CFG