Q UI C KIE PU LS E
The product is intended for use in the electromagnetic environment specified below. The user of the product should assure that it is used in
such an environment.
Immunity Test
IEC 60601 Test Level
TABLE 3
Conducted RF
20 Vrms
IEC 61000 4 6
150 kHz to 80 MHz
Outside ISM bands
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Radiated RF
20 V/m
IEC 61000 4 3
80 MHz to 2.5 GHz
NOTE 1 - At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 - These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from struc-
tures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF Transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the product is used, exceeds the applicable RF compliance level above, the product should be observed to verify nor-
mal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the
product.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 20 V/m.
Electromagnetic Environment Guidance
Compliance Level
20 V/m
Portable and mobile RF communications equipment should be used no
closer to any part of the product, including cables, than the
recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
20 V/m
Recommended separation distance:
d=1,2 P
d=1,2 P
d=1,2 P
where P is the maximum output power rating of the transmitter in
Watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey, should be less than the compliance level in
each frequency range.
Interference may occur in the vicinity of equipment marked with the
following symbol:
41
80 MHz to 800 MHz
800 MHz to 2.5 GHz
APPENDIX
b